11.2  Power failure alarm 

 

If at any time the mains power should be removed from the power unit or the power cable is 

unplugged without turning the power unit OFF, an audible alarm will be heard and the 

Power Failure

 

indicator will illuminate. The audible alarm may be cancelled by pressing the 

Alarm Mute

 button. 

 

 

 

Note: When the power unit has not been used for more than 3 months, it might 
require 60 minutes operating time (or more) for the Alarm to function correctly.

 

 

11.3  Service (alternating failure alarm) 

 

Should your system develop a fault condition whilst in use, an audible alarm will be heard and the 

Service

 indicator will illuminate. Please refer to 

Table 1

 for error codes and call Ultimate Healthcare 

or your local distributor.

  

 
 

11.4  Annual Service Indicator 

 

A dedicated 

Annual Service 

indicator on the control panel will illuminate and stay ON to alert nursing 

staff of the need for the system to be professionally serviced.  

 

The 

Annual Service 

indicator will illuminate after 365 days of use and the indicator light can only be 

extinguished by a professional service technician once the required service has been conducted. 

 
 
Table 1: Alert/Error Code Reference Table 
 

Priority 

High 

 

Low 

Warning 

Code 

Indicator 

LED 

Audible 

Output Mode 

Condition of 

Output 

Warning Description 

Remarks 

N/A

 

N/A 

ONCE 

Not in System 

Shutdown 

Key Tone 

Key Tone from  

Functional Button 

 

Power 

Failure 

ONCE 

Power-Off 

System Shutdown 

Shutdown 

 

ALL LED 

ONCE 

Operate or 

Standby 

Power-On 

All Indicators  

Light On 

N/A

 

N/A 

ONCE 

Operate or 

Standby 

State / Mode  

Switching 

No Display 

 

Auto-Firm 

ONCE 

Operate 

Mattress Inflation 

Completion 

Inflation Ended 

 

Auto-Firm 

ONCE 

Operate 

Auto-Firm Completion 

Auto-Firm Ended 

 

Static 

ONCE 

Operate 

Static Completion 

Static Ended 

N/A

 

Power 

Failure 

REPEAT 

(Cycle 4 sec.)

 

Power-Off 

Power Failure Alarm 

No Display 

 

Low 

Pressure 

REPEAT 

(Cycle 4 sec)

 

Operate or 

Standby 

Power-On Inflation 

Failure Alarm 

Inflate Failure 

 

Low 

Pressure 

REPEAT 

(Cycle 4 sec)

 

Operate or 

Standby 

Auto-Firm  

Failure Alarm 

Auto-Firm Failure 

Содержание UPRA3678D-2

Страница 1: ...Heritage II Digital Dynamic Mattress Replacement System UPRA3678D 2 User Manual Issue 4 16 08 18...

Страница 2: ...Page 2 of 29 Issue 4 16 08 18...

Страница 3: ...12 8 2 Comfort level 13 8 3 Auto detection 13 8 4 Control panel lockout 13 8 5 Function mode switch 14 8 5 1 Alternate 14 8 5 2 Static 14 8 6 Auto Firm 14 8 7 Alternate cycle time selection 14 8 8 Al...

Страница 4: ...Page 4 of 29 Issue 4 16 08 18 15 2 Fuse replacement 21 15 3 Air filter replacement 21 16 EMC Information 22...

Страница 5: ...tone 2 1 Contraindications The mattress is not intended to be used by anyone under 12 years of age The mattress is not suitable for use on patients with unstable fractures 3 About the Product The Heri...

Страница 6: ...xings Air Pipe Connector 3 x Static Head Cells 11 x Standard Cells Removable Cover 4 x Smaller Lower Legs Cells TPU 1 x Standard Cell Knitted Cell Retention Bands FOOT END HEAD END Air Pipe Connection...

Страница 7: ...hat requires correct procedures or practises in order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The...

Страница 8: ...power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suitably stowed before moving cleaning or mai...

Страница 9: ...p Classification Class II Type BF Warranty 2 years Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa W...

Страница 10: ...foot end as viewed from the foot of the bed There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient points on the bedframe patient surface Ca...

Страница 11: ...or to the power unit and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Страница 12: ...mfort Control Buttons 7 Operate Standby Button Indicators 3 Comfort Setting Indicators 8 Cycle Time Selection Button Alternate 4 Control Panel Lockout Button Indicator 9 Cycle Time Display 5 Function...

Страница 13: ...Auto Detection Once activated the indicator will flash and the system will start detecting and will set the internal pressure appropriately The indicator light will flash until the patient s weight an...

Страница 14: ...irm button In this mode all cells inflate to a single pressure setting to provide a firm and stable surface for nursing procedures or for patient ingress egress The system will automatically return to...

Страница 15: ...sures in the mattress will now equalise but maintain a degree of comfort This will maintain the cells in their present state for approximately 48 hours It is important to restore the Heritage II Digit...

Страница 16: ...ystem to be professionally serviced The Annual Service indicator will illuminate after 365 days of use and the indicator light can only be extinguished by a professional service technician once the re...

Страница 17: ...connected to the mains power supply and that the mains switch is turned ON Check power unit for any blown fuses Low Pressure indicator is flashing and sounding Check to ensure that the CPR tag is sec...

Страница 18: ...s resting against a solid surface Service indicator remains illuminated Indicates that system is scheduled for an annual service Please contact Ultimate Healthcare or your local dealer to arrange If t...

Страница 19: ...disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or disinfectants and not observing...

Страница 20: ...e dried or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceedin...

Страница 21: ...d plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any kink or...

Страница 22: ...RF emissions CISPR 11 Group 1 The GD311 401 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equi...

Страница 23: ...e s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short 5 UT 95 dip in 5 UT 95 dip in Mains p...

Страница 24: ...fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked w...

Страница 25: ...r W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 0 1 0 1 2 0 1 2 0 2 3 0 1 0 38 0 3 8 0 7 3 1 1 2 1 2 2...

Страница 26: ...Page 26 of 29 Issue 4 16 08 18...

Страница 27: ...Page 27 of 29 Issue 4 16 08 18...

Страница 28: ...Page 28 of 29 Issue 4 16 08 18...

Страница 29: ...sue 4 16 08 18 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimateh...

Отзывы: