11.2 Power failure alarm
If at any time the mains power should be removed from the power unit or the power cable is
unplugged without turning the power unit OFF, an audible alarm will be heard and the
Power Failure
indicator will illuminate. The audible alarm may be cancelled by pressing the
Alarm Mute
button.
Note: When the power unit has not been used for more than 3 days, it might require
30 minutes operating time (or more) for the Alarm to function correctly.
11.3 Service (alternating failure alarm)
Should your system develop a fault condition whilst in use, an audible alarm will be heard and the
Service indicator will illuminate. The audible alarm may be cancelled by pressing the
Alarm Mute
button. Should the situation not be resolved and the fault condition continues the alarm will resume.
Please contact Ultimate Healthcare or your local service provider.
12. Troubleshooting
PROBLEM
SOLUTION
No lights on power unit
Check the power unit is connected to the mains power
supply and that the mains switch is turned ON.
Check power unit for any blown fuses.
Low Pressure indicator is flashing
and sounding
Check to ensure that the CPR tag is securely fitted in
place.
Check whether power was suddenly shut down.
Check that the connection between air tube and power
unit is tightly secured.
Check that all tubing connections along the mattress
are secured.
If all of above steps have been checked. Press “Alarm
Mute” for system to be verified again.
Power Failure Alarm Indicator is
flashing and sounding
Check the power unit is connected to the mains power
supply and that the mains switch is turned ON.
Patient is bottoming out (without
alarm being triggered)
Pressure setting might be inadequate for the patient,
adjust comfort level to Firm and wait for a few minutes
for a better comfort.
The power unit is operating but the
mattress is not alternating
Ensure that the mattress inflation process is complete.
Check that t
he ‘Alternate’ indicator on the control panel
is illuminated. If not, press Function Select Button to
switch to alternating mode.
Power unit is noisy
Ensure that the power unit is resting against a solid
surface.
If the problem persists, contact Ultimate Healthcare or your local service provider.
Содержание Tamora Plus II
Страница 2: ...Page 2 of 24 Issue 5 13 02 2019...
Страница 18: ...Page 18 of 24 Issue 5 13 02 2019 16 EMC Information...
Страница 19: ...Page 19 of 24 Issue 5 13 02 2019...
Страница 20: ...Page 20 of 24 Issue 5 13 02 2019...
Страница 21: ...Page 21 of 24 Issue 5 13 02 2019...
Страница 22: ...Page 22 of 24 Issue 5 13 02 2019 Notes...
Страница 23: ...Page 23 of 24 Issue 5 13 02 2019 Notes...