13. Troubleshooting
PROBLEM
SOLUTION
No lights on Pump
Check the Pump is connected to the mains power
supply and that the mains switch is turned ON.
Check Pump for any blown fuses.
Low Pressure indicator is flashing
and sounding
Check to ensure that the CPR valve is securely closed.
Check whether power was suddenly shut down.
Check that the connection between air tube and Pump
is tightly secured.
Check that all tubing connections along the mattress
are secured.
If all of above steps have been checked. Press “Alarm
Mute” for system to be verified again.
Power Failure Alarm Indicator is
flashing and sounding
Check the Pump is connected to the mains power
supply and that the mains switch is turned ON.
The Pump is operating but the
mattress is not alternating
Ensure that the mattress inflation process is complete.
Check that the ‘Alternate’ indicator on the control panel
is illuminated. If not, press Function Select Button to
switch to Alternating mode.
Patient is bottoming out (without
alarm being triggered)
Pressure setting might be inadequate for the patient,
adjust comfort level by pressing Firm to the next weight
parameter, allow a few minutes for the mattress to
adjust the pressure and check again as outlined in 9.2
to ensure the patient does not continue to bottom out.
Pump is noisy
Ensure that the Pump is resting against a solid surface.
Service indicator remains
illuminated
Indicates that system is scheduled for an annual
service. Please contact Ultimate Healthcare or your
local service provider to arrange.
If the problem persists, contact Ultimate Healthcare or your local service provider.