5. Important Safety Information 

 
Please read all instructions prior to using any Ultimate Healthcare supplied product. The Heritage II 
Bariatric dynamic mattress replacement system must be used in accordance with this User Manual. 
 

 

The mattress must only be operated by personnel who have been properly trained or have 

suitable experience with products of this nature. 

 

Ensure a clinical Risk Assessment is conducted, which should take account of the suitability of 

use of this product, patient’s condition, any ancillary equipment in use and the surrounding 

environment. This should include assessing the use of side rails, head and footboards etc. 

Pressure settings should be advised or prescribed by a medical practitioner. 

 

Only personnel trained or formally approved by Ultimate Healthcare in operation and 

maintenance of Ultimate Healthcare products may perform maintenance; modification or repair 

work on any Ultimate Healthcare supplied product. 

 

Ensure the power cable is not trapped or twisted and is routed suitably to avoid crushing or 

entrapment when connected to the product. 

 

Do not use your mattress system power unit in the presence of flammable gases. This excludes 

oxygen cylinders. 

 

Avoid hazards caused by inappropriate handling of the power cable e.g. by kinking, shearing or 

other mechanical damages. 

 

The power cable for this product must be unplugged from the mains power outlet socket and 

suitably stowed before moving, cleaning or maintenance activities. 

 

Do not secure mattress straps to removable head or footboards or any fixed (non-moving) parts 

of a profiling bedframe.

 

 

Disconnect from mains (power supply) before cleaning the power unit. 

 

When cleaning do not immerse the power unit in water.

 

 

Use only the cleaning and disinfectant agents recommended in this User Manual. 

 

When connecting product after transportation or storage, inspect the power cable visually for any 

signs of damage. If evident, do not use product and contact Ultimate Healthcare or your local 

distributor for repair.

 

 

When storing, ensure the product is stored away from direct sunlight and extreme cold conditions. 

 

Never block the air openings of this product or place it on a soft surface, such as a bed or couch, 

where their openings may be blocked. Keep the air opening free of lint, hair, and other similar 

particles. Air ventilation through the power unit is vital for correct and safe operation. 

 

If this product is used for any activity other than detailed within this User Manual then personal 

risk to the end user or patient may occur. Ultimate Healthcare shall not be held liable/responsible 

for such an event. 

 

Do not smoke, or allow the patient to smoke when using this product. Keep all possible ignition 

sources clear of this product 

Содержание Heritage II Bariatric

Страница 1: ...Heritage II Bariatric Bariatric Dynamic Mattress Replacement System UPRA4878 2 User Manual Issue 5 12 12 2018 UPRA4878 2 UM 5...

Страница 2: ...Page 2 of 28 Issue 5 12 12 2018...

Страница 3: ...8 2 Comfort level 13 8 3 Auto detection 13 8 4 Control panel lockout 13 8 5 Function mode switch 14 8 5 1 Alternate 14 8 5 2 Static 14 8 5 3 Constant Low Pressure CLP 14 8 6 Auto Firm 14 8 7 Alternat...

Страница 4: ...Page 4 of 28 Issue 5 12 12 2018 16 Service and Maintenance 21 16 1 General 21 16 2 Fuse replacement 21 16 3 Air filter replacement 21 17 EMC Information 22...

Страница 5: ...anyone under 12 years of age The mattress is not suitable for use on patients with unstable fractures The mattress is not intended for use by patients with a BMI Body Mass Index of less than 30 3 Abou...

Страница 6: ...essure redistribution Mattress Replacement Power Unit CPR Tag Cable Management Fixings Air Pipe Connector 3 x Static Head Cells TPU 11 x Standard Cells TPU Removable Cover 4 x Smaller Lower Legs Cells...

Страница 7: ...that requires correct procedures or practises in order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The...

Страница 8: ...he power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suitably stowed before moving cleaning or m...

Страница 9: ...A Classification Class II Type BF Warranty 2 years Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa W...

Страница 10: ...d foot end as viewed from the foot of the bed There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient points on the bedframe patient surface C...

Страница 11: ...tor to the power unit and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Страница 12: ...ction 6 Auto Firm Indicator 2 Comfort Control Buttons 7 Operate Standby Button Indicators 3 Comfort Setting Indicators 8 Cycle Time Selection Button Alternate 4 Control Panel Lockout Button Indicator...

Страница 13: ...Once activated the indicator will flash and the system will start detecting and will set the internal pressure for the appropriately The indicator light will flash until the patient s weight and posi...

Страница 14: ...mattress will be reduced by 50 and the mattress will provide a comfortable static surface 8 6 Auto Firm Auto Firm mode can be selected by pressing the Auto Firm button In this mode all cells inflate t...

Страница 15: ...click into the OPEN position At the same time disconnect the air pipe connector from the power unit to speed up the air release If re inflating the mattress ensure the dial of the CPR attachment is ro...

Страница 16: ...le is unplugged without turning the power unit OFF an audible alarm will be heard and the Power Failure indicator will illuminate The audible alarm may be cancelled by pressing the Alarm Mute button N...

Страница 17: ...e 4 sec Operate or Standby Power On Inflation Failure Alarm Inflate Failure 9 Low Pressure REPEAT Cycle 4 sec Operate or Standby Auto Firm Failure Alarm Auto Firm Failure 10 Low Pressure REPEAT Cycle...

Страница 18: ...steps have been checked Press Alarm Mute for system to be verified again Power Failure Alarm Indicator is flashing and sounding Check the power unit is connected to the mains power supply and that the...

Страница 19: ...disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or disinfectants and not observing...

Страница 20: ...le dried or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceedi...

Страница 21: ...plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any kink or...

Страница 22: ...RF emissions CISPR 11 Group 1 The GD311 401 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equ...

Страница 23: ...ne s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short 5 UT 95 dip in 5 UT 95 dip in Mains...

Страница 24: ...m fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked...

Страница 25: ...er W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 0 1 0 1 2 0 1 2 0 2 3 0 1 0 38 0 3 8 0 7 3 1 1 2 1 2 2...

Страница 26: ...Page 26 of 28 Issue 5 12 12 2018...

Страница 27: ...Page 27 of 28 Issue 5 12 12 2018...

Страница 28: ...ue 5 12 12 2018 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Отзывы: