Ultimate Healthcare Heritage II Bariatric Скачать руководство пользователя страница 1

 

 
 

 

Heritage II Bariatric

 

Bariatric Dynamic Mattress Replacement System 

UPRA4878-2

 

User Manual 

Issue 5: 12/12/2018 

UPRA4878-2.UM-5 

 
 

Содержание Heritage II Bariatric

Страница 1: ...Heritage II Bariatric Bariatric Dynamic Mattress Replacement System UPRA4878 2 User Manual Issue 5 12 12 2018 UPRA4878 2 UM 5...

Страница 2: ...Page 2 of 28 Issue 5 12 12 2018...

Страница 3: ...8 2 Comfort level 13 8 3 Auto detection 13 8 4 Control panel lockout 13 8 5 Function mode switch 14 8 5 1 Alternate 14 8 5 2 Static 14 8 5 3 Constant Low Pressure CLP 14 8 6 Auto Firm 14 8 7 Alternat...

Страница 4: ...Page 4 of 28 Issue 5 12 12 2018 16 Service and Maintenance 21 16 1 General 21 16 2 Fuse replacement 21 16 3 Air filter replacement 21 17 EMC Information 22...

Страница 5: ...anyone under 12 years of age The mattress is not suitable for use on patients with unstable fractures The mattress is not intended for use by patients with a BMI Body Mass Index of less than 30 3 Abou...

Страница 6: ...essure redistribution Mattress Replacement Power Unit CPR Tag Cable Management Fixings Air Pipe Connector 3 x Static Head Cells TPU 11 x Standard Cells TPU Removable Cover 4 x Smaller Lower Legs Cells...

Страница 7: ...that requires correct procedures or practises in order to prevent personal injury BF symbol indicates this product is according to degree of protection against electric shock for type BF equipment The...

Страница 8: ...he power cable e g by kinking shearing or other mechanical damages The power cable for this product must be unplugged from the mains power outlet socket and suitably stowed before moving cleaning or m...

Страница 9: ...A Classification Class II Type BF Warranty 2 years Operation environment 50C to 400C 15 RH 93 RH no condensation Storage environment 25 70 93 RH no condensation Environment pressure 70 kPa 101 3 kPa W...

Страница 10: ...d foot end as viewed from the foot of the bed There are security straps fitted to the base cover of the mattress which should be fastened loosely to convenient points on the bedframe patient surface C...

Страница 11: ...tor to the power unit and then ensure that the air tubes are free from any kind of obstruction and are not kinked Plug the power cable into a suitable electrical socket and switch ON using mains power...

Страница 12: ...ction 6 Auto Firm Indicator 2 Comfort Control Buttons 7 Operate Standby Button Indicators 3 Comfort Setting Indicators 8 Cycle Time Selection Button Alternate 4 Control Panel Lockout Button Indicator...

Страница 13: ...Once activated the indicator will flash and the system will start detecting and will set the internal pressure for the appropriately The indicator light will flash until the patient s weight and posi...

Страница 14: ...mattress will be reduced by 50 and the mattress will provide a comfortable static surface 8 6 Auto Firm Auto Firm mode can be selected by pressing the Auto Firm button In this mode all cells inflate t...

Страница 15: ...click into the OPEN position At the same time disconnect the air pipe connector from the power unit to speed up the air release If re inflating the mattress ensure the dial of the CPR attachment is ro...

Страница 16: ...le is unplugged without turning the power unit OFF an audible alarm will be heard and the Power Failure indicator will illuminate The audible alarm may be cancelled by pressing the Alarm Mute button N...

Страница 17: ...e 4 sec Operate or Standby Power On Inflation Failure Alarm Inflate Failure 9 Low Pressure REPEAT Cycle 4 sec Operate or Standby Auto Firm Failure Alarm Auto Firm Failure 10 Low Pressure REPEAT Cycle...

Страница 18: ...steps have been checked Press Alarm Mute for system to be verified again Power Failure Alarm Indicator is flashing and sounding Check the power unit is connected to the mains power supply and that the...

Страница 19: ...disinfecting ensure that only a damp cloth is used Caution After cleaning dry the mattress out of direct exposure to sunlight Caution Using inappropriate detergents or disinfectants and not observing...

Страница 20: ...le dried or air dried They may be tumble dried on a low heat for up to 90 minutes Drying temperature must not exceed 400C The mattress cover and mattress must be dry prior to refitting Caution Exceedi...

Страница 21: ...plug if there are abrasions or excessive wears 2 Check mattress cover for signs of wear or damage Ensure mattress cover and tubes are stubbed together correctly 3 Check the air hoses for any kink or...

Страница 22: ...RF emissions CISPR 11 Group 1 The GD311 401 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equ...

Страница 23: ...ne s 2kV line s to earth 1kV differential mode Not applicable Mains power quality should be that of a typical commercial or hospital environment Voltage Dips short 5 UT 95 dip in 5 UT 95 dip in Mains...

Страница 24: ...m fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked...

Страница 25: ...er W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 0 1 0 1 2 0 1 2 0 2 3 0 1 0 38 0 3 8 0 7 3 1 1 2 1 2 2...

Страница 26: ...Page 26 of 28 Issue 5 12 12 2018...

Страница 27: ...Page 27 of 28 Issue 5 12 12 2018...

Страница 28: ...ue 5 12 12 2018 Ultimate Healthcare Ltd Calmore Industrial Estate Nutwood Way Totton Southampton Hampshire SO40 3WW Tel 0333 321 8996 Fax 023 8066 2388 Email info ultimatehealthcare co uk www ultimate...

Отзывы: