Troubleshooting
OTHER PROBLEMS WHICH MAY OCCUR
ACTION
A test strip has not been inserted into the
meter properly.
Review the instructions and reinsert a
test strip correctly.
Defective test strip.
Replace with a new test strip.
A test strip remains in the test port for more than
2 minutes before testing.
Meter will automatically turn off. Reinsert a test strip
into the test port.
LCD display on the meter is blank when trying to perform
a test.
Check batteries and if problem persists contact
customer service for assistance.
The meter does not function after new batteries
are installed.
Please remove batteries and wait for 1 minute, then
reinstall batteries to see if meter is working properly.
Be sure batteries are installed correctly.
Specifications
Working Temperature Humidity
50°F - 104°F (10°C - 40°C)
Hematocrit
10 - 90% RH
Test Sample
20 - 60%
Sample Volume
0.7 µl
Measuring Unit
mg/dL or mmol/L
Measuring Range
20 - 600 mg/dL (1.1 - 33.3 mmol/L)
Test Time
5 seconds countdown
Memory Capacity
500 most recent results
Average
7/14/30 days average
Power Supply
2 AAA batteries
Battery Life
Approximately 1000 tests
Dimensions
106L x 66W x 20H mm
Weight
65 g without batteries
Meter Storage
Store at temperature -4°F - 122°F (-20°C - 50°C), Less than 90% RH