5)
The skin must be shaved before treatment.
6)
The handle piece is better to be operated at
SCANNING
mode. It is
faster for treatment.
7)
Skin contact gel is no necessary for treatment. It won’t make the
treatment better or improve the skin feeling.
8)
You need to wait for several minutes before the cooling tip becomes cold.
9)
Keep contact between the cooling tip and skin. And you can also use the
edge of the Aluminum head of the handle piece to cool down the skin
during treatment.
10)
Never shot the laser pulses to eyes, or watch the laser directly.
Chapter 5: Maintenance and Warranty
This chapter descript the daily maintenance including cleaning the system, cleaning the crystal.
Maintenance
1)
The water must be checked and changed every 6 months.
2)
The machine must be operated in a room with air-conditioner. The room temperature is
better <28°C.
3)
The environment temperature can not go below ZERO, or the ice may damage the water
tanks and water channels inside the machine and the handle piece.
4)
Clear the gel on the handle piece and make sure there is no dust attached on the handle piece
window.
5)
Do not bend the tube of the handle piece too hard, or this may cause flow switch ERROR.
6)
Do not plug in & out the handle piece connectors frequently. This may cause bad
connections and may cause water leakage or not well electrical connections.
7)
Do not shot the laser at random targets, or it may cause burning or fire.
Warranty Notice
Please take necessary protections for the operation of the machine and handle piece, and notice
that following situations can not be warranted:
1)
Operation the system without water.