5 Maintenance

While any service or repairs of the instrument are performed by Ugo Basile, this section 

describes normal maintenance procedures that can be performed at your facility.

UNPLUG THE POWER CORD BEFORE CARRYING OUT ANY MAINTENANCE WORK

5.1 Electrical

To inspect and/or replace the fuses, disconnect the mains cable first! Insert a miniature screw-

driver in the slot indentation and snap out the slide which houses the fuses. Use T400mA fuses 

for operation at both 230 or 115 Volts.

Snap in the fuse slide: the mechanical “click” ensures that it is locked.

5.2 Cleaning/disinfection

The Librae unit (47880) does not require any maintenance apart from normal cleaning.

Do not use organic solutions on the restrainer and cabinet, as they are liable to impair the 

transparency of the restrainer and crack the acrylic component and the touch pad. 

Cotton wool and water can be used for cleaning purposes. To disinfect, use a non-alcoholic 

disinfectant, or H

2

O. 

5.3 Long Inactivity

The instrument does not require any particular maintenance after long inactivity, except 

cleaning

5.4 Customer Support

For any further information you may require concerning the use and/or maintenance of the 

Librae Incapacitance Tester, please do not hesitate to contact our service department (or our 

local distributor) either directly of via our support page. Before sending any instrument to 

our factory for repair, please contact our logistics department to obtain a return authorization 

number (RMA) and shipping/packing instructions. We may not be held responsible for damages 

caused during transport due to poor packing. Whenever possible, please use the original 

packing.

6 Specifications

General

Command Input

4.3’’ touch-screen (resistive)

Read-out

4.3’’ TFT touch-screen

Power Requirements

Universal input 85-264 VAC, 50-60Hz, 15W max.

Sound Level

< 45dB(A)

Operating Environment

10°C to 40°C; 5% to 95% RH (non-condensing)

Operation

Animal weight (per paw)

from 20g to 4400g

Precision

from 0.05% to 0.1%

Measurement Duration

from 1s to 360s

Measurement Mode

Manual and Autostart

Autostart parameters

Weight threshold, Stabilization Threshold, 

Sustain Time

Weight Averaging

1 sample to 10 samples

Start Experiment

Start button, pedal switch, or Autostart

Stop Experiment

Stop button, pedal switch, or elapsed  

timeout

Data export

.csv format, from USB key (provided)

TTL Output

Start/Stop, Measurement Mode.

5VDC +/- 10%; 10mA MAX

Physical

Total Weight

2.5kg

Shipping Weight

5kg

Dimensions (without restrainer)

25cm(w) x 31cm(d) x 13cm(h)

Dimensions (including restrainer)

25cm(w) x 31cm(d) x 25cm(h)

Packaging dimensions

68cm(w) x 34cm(d) x 28cm(h)

Warranty

Product Warranty

The Librae Incapacitance Tester unit (47880) is 

covered by a 12-month warranty. Coverage can 

be increased to 24-months for free when you 

register your instrument on the UGO BASILE 

website.

Содержание 47883

Страница 1: ...Instruction manual Librae Incapacitance tester July 2020 revision 2 2 SKU 47883 47882 47885 47880 001 Pain Inflammation...

Страница 2: ...organization Instruction Manual July 2020 Revision 2 2 MOD 13 Rev 1 CE CONFORMITY STATEMENT Manufacturer UGO BASILE srl Address Via G di Vittorio 2 21036 Gemonio VA ITALY Phone n 39 0332 744574 Fax n...

Страница 3: ...d of this manual and also at ugobasile com Product features What is in the box Librae package contents SKU 47883 The Librae Incapacitance tester for mice 1 transparent plastic restrainer for mice 1 Pe...

Страница 4: ...e the weight averaged over a few seconds is normally used as the data point For example the device provides an index of discomfort in the osteoarthritic knee normally expressed as the Left Right weigh...

Страница 5: ...th the Librae touch screen you may notice a different less sensitive response You may need to press slightly harder or use a stylus pen on the touch screen This is because we employ a resistive touch...

Страница 6: ...e you are welcome to contact our Service Department see paragraph 5 4 Customer Support specifying the serial number of your instrument 3 3 Safety Instructions The following guidelines must be followed...

Страница 7: ...e manufacturer Servicing carried out by unauthorized personnel See also paragraphs 2 1 1 and 2 1 2 3 8 Connection Connect the mains cord between the power socket of the Librae and the power outlet pro...

Страница 8: ...might take up to 60 seconds During this time the UB logo and product name will be displayed Figure 4 0 1 Figure 4 0 1 4 1 Home Page When the start up is finished the Home Page Menu is displayed Figur...

Страница 9: ...l has entered the restrainer tilt the restrainer back down to the flat position 8 When ready press START button or pedal switch to run the experiment 6 1 If Autostart Mode is enabled wait for the inst...

Страница 10: ...text field to enable a virtual keyboard Figure 4 2 2 This is an alphanumeric field enter a maximum of 12 characters Figure 4 2 2 Characters Virtual Keyboard Figure 4 2 3 Numeric Virtual Keyboard PROT...

Страница 11: ...the restrainer and if the experiment is running correctly Figure 4 4 1 Ehorizontal lements displayed on the START page from left to right provide the following functionality 1 i INFO button This butto...

Страница 12: ...start Autostart is a unique Librae Incapacitance Tester feature By selecting this operative mode Librae helps the user to perform the experiment in a consistent way reducing variability and user bias...

Страница 13: ...tabilization AUTO icon is YELLOW When ALL the autostart requirements are completed the experiment automatically starts Figure 4 6 5 Figure 4 6 5 Autostart is waiting for stabilization time to elapse A...

Страница 14: ...or update system icon refer to paragraph 4 1 1 5 Press Update button and wait for a reboot Figure 4 7 1 1 Now the Librae Incapacitance Tester is up to date An Erase DB could be necessary Refer to para...

Страница 15: ...ton shows Figure 4 7 5 1 the results of the experiments stored inside Librae Incapacitance Tester Every result box contains 1 Date and time of the experiment 2 Treatment ID protocol stage and trial 3...

Страница 16: ...gure 4 8 1 1 and Table 4 8 1 1 for connector pinout Figure 4 8 1 1 DB 15 Pin Signal Name Signal Type Description 1 Start Stop TTL OUT Start TTL High Level Stop TTL Low Level 2 Manual Autostart TTL OUT...

Страница 17: ...not hesitate to contact our service department or our local distributor either directly of via our support page Before sending any instrument to our factory for repair please contact our logistics dep...

Страница 18: ...paratus Thermal Place Preference TPP Test for Mice Rats Analgesy Meter the 1st and original Randall Selitto paw pressure test Other Ugo Basile products related to this device visit ugobasile com web s...

Страница 19: ...torio 2 21036 Gemonio VA ITALY Tel 39 0332 744574 sales ugobasile com The 1st original Mouse RotaRod for motory coordination studies Fear Conditioning System ANYmaze GSM Grip Strength Meter for mice a...

Отзывы: