47650 Instruction Manual (Rev. 2.1)

 

  

 

3.5.1 

Mains Switch 

This two-pole toggle switch, which complies with international safety standards, is lighted 
when the instrument is 

ON

3.5.2 

Fuse Holder  

The fuse holder comprises two fuses, one on the live, and the other on the neutral.  Use 
(T2A) timed fuses for operation at both 115 or 230 Volts.  

For fuse replacement, please refer to paragraph 6.1-Electrical.  

3.5.3 

Mains Cord  

It is a standard cable, Cat. # E-WP008. Make sure your power outtake is provided with a 
reliable ground connection. 

3.6 

Intended Use 

The Rota-Rod is intended for investigation use on 

laboratory animals

 

only

3.7 

Additional Safety Consideration 

1) 

Use  original  accessories  and  spare  parts  only,  see  also  paragraph  8-
ORDERING INFORMATION.
 

2) 

immediately disconnect and replace damaged mains cord. 

3) 

do not obstruct a comfortable access to the power module. 

4) 

do  not  operate  in  hazardous  environments  or  outside  prescribed 
environmental limitations (i.e. +10c° / +40c°, 95% max. relative humidity, non-
condensing), see also paragraph 7-47650 SPECIFICATIONS. 

5) 

do not spray any liquid on the connectors and on the geared motor. 

UGO  BASILE  DOES  NOT  ACCEPT  ANY  RESPONSIBILITY  FOR  PROBLEMS  OR 
HARM CAUSED TO THINGS OR PERSONS, ARISING FROM: 

 

incorrect electrical supply; 

 

incorrect installation procedure; 

 

incorrect  or  improper  use  or,  in  any  case,  not  in  accordance  with  the 
purpose  for  which  the  instrument  has  been  designed  and  the  warnings 
stated in the instruction manual supplied with the instrument; 

 

replacement  of  original  components,  accessories  or  parts  with  others  not 
approved by the manufacturer; 

 

servicing carried out by unauthorized personnel.  

Содержание 47650

Страница 1: ...UGO BASILE S R L sales ugobasile com service ugobasile com www ugobasile com instruction manual Rotarod Mouse Cat No 47650 E C T Unit Cat No 57800...

Страница 2: ......

Страница 3: ...ED TO ENSURE SAFE OPERATION AND TO RETAIN THE INSTRUMENT IN SAFE CONDITIONS SERVICE AND ADJUSTMENTS SHOULD BE CARRIED OUT BY QUALIFIED PERSONNEL AUTHORIZED BY UGO BASILE ORGANIZATION ANY ADJUSTMENT MA...

Страница 4: ...TION MANUAL MANUALE DI ISTRUZIONE X PAD SOFTWARE SOFTWARE DATE Serial No IMBALLATO DA PACKED BY Set for 115V 50 Hz 230V 50 Hz 115V 60 Hz 230V 60 Hz IMPORTANT IMPORTANTE Check the shipment for complete...

Страница 5: ...PPLYING POWER 5 3 5 1 MAINS SWITCH 6 3 5 2 FUSE HOLDER 6 3 5 3 MAINS CORD 6 3 6 INTENDED USE 6 3 7 ADDITIONAL SAFETY CONSIDERATION 6 3 8 CONNECTIONS 7 3 8 1 USB HARD DISK 7 3 8 2 KEYBOARD 7 3 9 INSTAL...

Страница 6: ...D 22 5 8 SIGNAL OUTPUT 22 5 8 1 PIN OUT 22 6 MAINTENANCE 23 6 1 ELECTRICAL 23 6 2 CLEANING 23 6 3 LONG INACTIVITY 24 6 4 CUSTOMER SUPPORT 24 7 47650 SPECIFICATIONS 24 8 ORDERING INFORMATION 25 8 1 OPT...

Страница 7: ...the trip box below the plate boxes and the corresponding magnetic switch is activated thereby recording the animal endurance time in seconds The display shows the actual angular speed RPM At the end o...

Страница 8: ...tes a 4 3 touch screen for basic setting and monitoring via an intuitive panel The picture shows the Rota Rod main menu Figure 1 Main Menu During the test the touch screen indicates the actual speed R...

Страница 9: ...is dependent on the conductive nature of human body it doesn t work if the user is wearing gloves 2 2 Rota Rod Features The main features of the new model are SPEED adjustable in the range 3 80 RPM in...

Страница 10: ...r in full version The software automatically classifies the data combining configuration settings with test results The user can add information before or after the test Results appear in a tree like...

Страница 11: ...ownload on our web For any additional information and or assistance you are welcome to contact our Service Department see paragraph 6 4 Customer Support specifying the serial number of your instrument...

Страница 12: ...raph 8 ORDERING INFORMATION 2 immediately disconnect and replace damaged mains cord 3 do not obstruct a comfortable access to the power module 4 do not operate in hazardous environments or outside pre...

Страница 13: ...ing and debug ETH the Ethernet connector will be used for remote diagnosis and Internet access used for standard operation 3 8 1 Hard Disk The device hard disk is now embedded in the inside electronic...

Страница 14: ...e one and requires some patience and gentle handling to be perfected Speaking about animal machine interface the surface finishing of the drums is a sensitive point The rotor drums are knurled to prov...

Страница 15: ...It advisable to carry out an acclimation and training session before starting the motor co ordination test With the rotor at low speed say 4 5 R P M each na ve mouse is placed upon its section in orde...

Страница 16: ...logo button always brings back to the Home page 5 1 1 System Icons NEW SOFTWARE DETECTED if this icon is present a new version of SW has been detected on the USB Key ETHERNET DETECTED an Ethernet conn...

Страница 17: ...Depress the keybord icon to activate the virtual keyboard Enter a maximum of 12 characters in each field TREATMENT PROTOCOL STAGE and TRIAL are text boxes where it possible to specify the related info...

Страница 18: ...ing the related button at the page bottom Having selected the Speed and mode press to go back to the Home page 5 3 1 Constant Speed With Constant Speed mode the rod rotates CCW at a speed which remain...

Страница 19: ...by a phase at decreasing speed a change of rotation hence a CW rotation phase at increasing speed followed by a phase at decreasing speed The scheme on the right simplifies the re verse cycle Depress...

Страница 20: ...then repeating the cycle until stopped number In the example the rotor will revolve repeatedly 5 times forward then 5 times backward Depress the icon to activate the numeric keyboard to enter Speed b...

Страница 21: ...2 Starting a Ramp In the accelerating mode when the RUN button is depressed the Rota Rod will pass from the IDLE state motor and counters disabled to the initial speed ALL Trip Boxes must be in the U...

Страница 22: ...cating the completed revolutions the data are saved and will appear in the RESULT table As soon as the FINAL SPEED is reached the rod wil keep rotating at constant speed and the timers will continue c...

Страница 23: ...and restart accelerating from INITIAL SPEED to FINAL SPEED The cycle is repeated until the STOP button is depressed at any time during the experiment When an animal falls off its rotor section tiltin...

Страница 24: ...ated information entered by the user appear under the blue headings DATE TIME LANE TREATMENT ID PROTOCOL STAGE TRIAL The test results appear under the red headings LATENCY REVOLUTIONS SPEED MODE DISTA...

Страница 25: ...at 5 6 2 Load Protocol When a USB flash drive containing at least one protocol is connected to the device the LOAD PROTOCOL button will open the following window displaying all the loadable protocols...

Страница 26: ...Rota Rod C Insert the USB KEY containing RR_Update hex file into one of the two USB ports If the USB key is detected the icon on the system bar will appear on the tool bar If the system recognizes a n...

Страница 27: ..._ 5 7 3 Diagnostics By selecting this function the Rota Rod will perform an self diagnostics automatic test First of all the machine tests the correct functioning of all trip boxes Follow the instruct...

Страница 28: ...ion and speed The 15 pin D SUB connector see picture and paragraph 3 8 provides different output trigger in the form of TTL signals All pins except pin 8 are TTL digital signals Pin 8 is an analog out...

Страница 29: ...E MAINS CORD BEFORE CARRYING OUT ANY MAINTENANCE JOB 6 1 Electrical To inspect and or replace the fuses disconnect the mains cable first Insert a miniature screwdriver in the slot indentation see para...

Страница 30: ...istics ugobasile com sales ugobasile com Before sending any instrument to our factory for repair please contact our logistics department to obtain a return authorization number RMA and shipping packin...

Страница 31: ...e Physical Total Weight 11Kg Shipping Weight 16Kg approx Dimensions 46 w x28 d x33 h cm Packing Dimensions 70x36x46cm Warranty 47650 is covered by a 24 month warranty 8 ORDERING INFORMATION 47650 Mous...

Страница 32: ...Motor Neuron Improves Neuromuscular Function and Motor Neuron Survivain Mutant SOD1 Mice Neurobiol Of Aging 35 4 906 915 2014 M Milanese et alia Knocking Down Metabotropic Glutamate Receptor 1 Improve...

Страница 33: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 34: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 35: ..._ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _...

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ...VALIDATION CERTIFICATE...

Отзывы: