Instructions for use (rev 290319)
ARTIS NEO
UGIN DENTAIRE
33
9.
MAINTENANCE AND PRECAUTIONS FOR USE
To maintenance work and/or cleaning, ARTIS NEO must be disconnected from the main
power supply, and at room temperature: wait the cooling after use (risk of burns and/or
damage)
Except for cleaning the machine outside, ARTIS NEO furnace maintenance entails only two other
operations:
-
Cleaning the platform sealing joint regularly with alcohol
-
Replacing the muffle after 3000 hours of service
Recommendation:
-
For press the control panel keys never use metal or very hot instruments which would
damage the plastic film
-
Clean the panel with a solvent-free product
-
Keep hot pieces away from the panel to avoid the risk of damage by heat radiation
-
Leave the furnace hooked up at all times, except during prolonged absence
-
Clean the machine outside with a soft, dry cloth
-
Do not introduce stranger objects inside the unit during the maintenance operations
-
Do not spill liquid on the ventilation openings or inside the furnace: risk of electric shock
9.1.
Decontamination program
Carry out regular decontamination program:
Set the program parameters as follows:
-
Pre-drying temperature = 700°C
-
Vacuum hold time
= 2 min
-
Platform rise time
= 0 min
-
Heat-rise rate
= 100°C/min
-
Internal pre-drying
= 0 min
-
Final temperature
= 1040°C
-
Vacuum start
= 700°C
-
Hold time
= 4 min
-
Vacuum stop
= 1030°C
-
Internal cooling
= 2 min
-
Vacuum level
= 94%
-
Platform lowering
= 0 min
Then place the "CARBONET" decontamination graphite on the base
Initiate the cycle by pressing the key