Instructions for use (rev 201221)
VACUUM PUMP
UGIN DENTAIRE
5
4.
INSTALLATION AND START-UP
4.1.
UNPACK
-
Unpack
VACUUM PUMP
and check that it is in perfect order.
-
Any defects can be reported to the carrier.
-
Remember to get the carrier to sign the delivery note.
NOTE: keep the original packaging to transport.
4.2.
INSTALLATION
-
For correct operation, the pump must be used in a well-ventilated room, without humidity, and with low
levels of dust. The room temperature should not exceed 40° C.
-
If it is not used, keep it in a room without humidity and protect it from dust and vibrations.
-
It is recommended that the device be levelled horizontally.
-
Keep around ventilation enough space to ensure good ventilation (at least 150 mm). Do not obstruct the
ventilation system for protection against overheating
ATTENTION: do not place the apparatus inside a piece of furniture without circulation of air.