Manual_500 ECO_SN0001-following_0520

500 ECO

www.ufsk-osys.com

User and Service Manual

Page 28 of 50

6.8. Reset function

The Reset key allows you to return the chair electronically into the factory-defined original patient entry and exit 
position (the „sitting position“) .

During this automatic movement, all drives will move into their pre-programmed initialization positions. 

The  Reset  function  serves  both  for  restarting  the  chair  in  case  of  a  malfunction  and  for  maintenance  of  the 

electronic system!

During all adjusting processes, the chair‘s control system continuously monitors the positions of the individual 

drives.  To  make  this  monitoring  possible,  the  control  system  initially  needs  to  know  drives‘  current  starting 

positions. In the event the control system loses a motor‘s position during operation (caused e.g. by a voltage 
interruption during a motor‘s movement or by the replacement of a motor), it will be necessary to re-initialize the 
control system.

 

Important Information

Resets conducted at regular intervals – we recommend  1-2 resets per week – will help to virtually rule out 
malfunctions of the electronic system.

The  following  situations  will  necessitate  an  initialization  as  per  Section  „6.8.2  Initialization  in  the  event  of  a 
malfunction“

 in case of a malfunction:

•  It is no longer possible to execute the individual adjusting functions at full speed, but only at half or moderate 

speed (”creep speed of the drives“) or the green AUTO LED lights steadily after a memory   position was 

reached.

•  the automatic movements are blocked
•  the memorization function is blocked

6.8.1 Routine reset for maintenance of the electronic system

6.8.1.1 Manual reset

It is possible to initialize the control system manually by keeping the Reset key permanently pushed until the green 

operator indicating LED of the manual keypad flashes twice in quick succession. The green operator indicating 

LED will light up steadily during the entire initialization movement.

6.8.1.2 Automatic reset

It is also possible to retrieve the chair’s desired initialization in the form of an automatic movement. To start the 

initialization movement, push the Auto key (1 second), followed by pushing of the Reset key. During the initialization 

movement, the green AUTO LED will flash. Completion of the automatic reset is indicated by double flash of the 

green operator indicating LED.

6.8.2 Initialization in the event of a malfunction

Failure of the chair to respond to the customary entries (blocked key commands) or stopping of the chair failure 
to execute the adjusting functions at full power (creep speed of the drives) may indicate the occurrence of an 
electronic control system error (los of motor position during adjusting) which can be eliminated by an initialization 

movement. A malfunction caused in this manner will be indicated by the green flashing of the AUTO LED and can 

be easily eliminated by the keystroke sequence of an automatic reset as described in 

Section „6.8.1.2 Automatic 

reset“. Completion of the initialization movement is indicated by double flash of the green operator indicating LED.

Содержание 500 ECO

Страница 1: ...GmbH Kirchhoffstrasse 1 93055 Regensburg Germany Phone 49 0 941 78862 15 Fax 49 0 941 78862 35 Mail info ufsk osys com Web ufsk osys com Version 05 2020 SN 0001 following User and Service Manual Trea...

Страница 2: ...without the written permission of the publisher or processed duplicated or distributed by means of electronic systems We accept no liability for minor technical changes color variations or typesetting...

Страница 3: ...on the signal words used herein 7 2 Safety 8 2 1 Basic safety instructions 8 2 2 Personnel Qualification 10 2 3 Intended Use 10 2 4 Improper Use 11 2 5 Equipment Labeling and Nameplate 11 3 Technical...

Страница 4: ...et memory position 27 6 8 Reset function 28 6 8 1 Routine reset for maintenance of the electronic system 28 6 8 1 1 Manual reset 28 6 8 1 2 Automatic reset 28 6 8 2 Initialization in the event of a ma...

Страница 5: ...ions 40 9 1 Weights and Measurements 40 9 2 Electrotechnical Data 41 9 3 Classification 41 9 4 Ambient conditions 41 10 Technical Drawing and Spare Parts List 42 10 1 Technical Drawing 42 10 2 Spare P...

Страница 6: ...ght to this user manual remains the property of UFSK International OSYS GmbH Regensburg This instruction and service manual is intended only for the operator and its staff It contains directions and i...

Страница 7: ...perate the treatment chair or to use it for conducting patient examinations and or administering patient treatments The user has been instructed in the handling of the treatment chair The user is capa...

Страница 8: ...g the treatment chair Please keep this instruction and service manual readily available at the treatment chair s location of use Please exercise caution when performing the works described in this use...

Страница 9: ...l adjustment functions is safely interrupted The same applies to the use of high frequency surgery equipment Electromagnetic interferences between these devices and the treatment chair cannot be ruled...

Страница 10: ...he treatment field head face allow an ergonomic working method especially optimized for ambulant operation techniques with high patient turnover IVOM treatments ophthalmology aesthetic treatments derm...

Страница 11: ...tion 7 8 9 11 6 10 3 4 5 1 2 Item Explanation of the symbols and or technical details 1 Data Matrix with UDI Unique Device Identification 2 UDI comprising 01 GTIN Global Trade Item Number 11 Date of M...

Страница 12: ...aracteristics Adjustment Patient entry OP position in 5 6 seconds Powerful operation via integrated power supply unit incl isolating transformer with protective low voltage Powerful electromechanical...

Страница 13: ...ent onto an I V Pole Side rails right left swiveling Padded arm board with universal joint Norm rail seat part Instrument table Paper roll holder stayClean roll paper roll Foot protection foil Trapeze...

Страница 14: ...essory mounting support powder coated with accessory holder stainless steel Item No 3351 1033 Accessory mounting support stainless steel Item No 3351 0037 Hand remote bracket Item No 3351 1023 Monitor...

Страница 15: ...5 cm hight adjustable swiveling detachable incl adapter Item No 3065 0000 Paper roll holder Item No 3351 1031 Paper rolls stayClean roll white embossed 2 ply 50 cm wide 9 rolls 50 m Item No 3034 0021...

Страница 16: ...00 mm wheel body in polyamide tread in polyurethane One double castor diameter 35 mm Wheel body made of polyamide Footstep lever Double step lever mechanical made of polyamide strengthened with glass...

Страница 17: ...f necessary please transport and store the treatment chair under the following conditions With all attachments disassembled In the chair s original packaging Secured against impacts and tipping In dry...

Страница 18: ...g In case of high temperature fluctuations during storage and shipping the chair should be given an acclimatization period of approx 2 hours to observe the maximum admissible temperature change of 20...

Страница 19: ...from pallet 6 Loosen the back section slide and remove the adhesive material from the back section 7 Loosen bolts for connecting the back drive and connecting shaft of the armrest attachment 8 Snap th...

Страница 20: ...d 14 move the backrest element back to approx 30 degrees 15 Position the seat cushion correctly 16 and snap into place by pressing firmly on the buttocks knee side 17 Move the chair into the sitting p...

Страница 21: ...elivered the mains connection cable can be replaced by a qualified electrician see Section 11 2 1 Replacing the mains connection cable 5 4 Switch on Connect the mains connection cable to an easily acc...

Страница 22: ...g and dismounting position upright sitting position The optionally available foot keypad can be used to activate the two free configurable memory positions 1 2 of the manual keypad This allows any alr...

Страница 23: ...ymbols Meaning explanation Reset button initialization restart the electronic control causes the chair to return to its defined home position Chair position mounting dismounting position for the patie...

Страница 24: ...tivates the automatic mode the control system waits for the in put command Select memo position 1 to 8 Auto Run LED lights green 8 Select memory position 1 to 8 and start it eight storage locations fo...

Страница 25: ...oot keypad can be positioned individually on the floor in the chair s environment 6 3 Switch on the chair To switch the electrically adjustable treatment chair on or off please push either the power k...

Страница 26: ...e patient s orthopedic handicaps when selecting the chair s adjusting functions to prevent bodily harm to the patient resulting from incorrect positioning and bedding Please make sure of adequate safe...

Страница 27: ...push of a selection key This conveniently obviates the need for the burdensome and time consuming activation of a series of individual adjusting functions Caution It is not permitted to activate the...

Страница 28: ...AUTO LED lights steadily after a memory position was reached the automatic movements are blocked the memorization function is blocked 6 8 1 Routine reset for maintenance of the electronic system 6 8 1...

Страница 29: ...g of the headrest element This ensures an optimally comfortable positioning of the patient through individually adjusted distances in the area of the head cervical spine support of the shoulders This...

Страница 30: ...eatment chair s left side and vice versa Armrest adjustment high Armrest adjustment low e g for children Caution Patients may only mount or dismount the treatment chair via the chair s large side The...

Страница 31: ...ed swivel castor 2 fixed castors rigid and 1 swivel castor raised retracted Treatment chair stands stable on 4 adjustable feet 2 fixed castors rigid and 1 swivel castor extended Treatment chair mobile...

Страница 32: ...treatment chair is intended for use with a direct mains connection The mains voltage is transformed into protective extra low voltage via the wide voltage power supply unit 100 240V integrated in the...

Страница 33: ...Plastic surfaces e g chassis become matt upon repeated application of too aggressive agents and also rough e g acetone high concentration of quaternary ammonium compounds Attention Electrical componen...

Страница 34: ...Manual_500 ECO_SN0001 following_0520 500 ECO www ufsk osys com User and Service Manual Page 34 of 50 Standard upholstery see skai Tundra electrically conductive upholstery see skai Mano S...

Страница 35: ...onsultation with the supplier will result in the automatic lapse of the manufacturer s warranty and all product liability The treatment chair has service life of approx ten 10 years subject to average...

Страница 36: ...le jackets abrasions bulges brittle areas visible internal insulation or metallic wires Damages to casings or coverings manual or foot keypad motor housing plugs Breaks ruptures cuts deformations abra...

Страница 37: ...SK International OSYS GmbH Accessories Medical device class I non sterile active electrically operated with measuring equipment x without measuring equipment no metrological test required Explosion pr...

Страница 38: ...justment Back seat and legrest Armrests Function check of manoeuvering aid Lowering of the castors Locking Function check of the electronics Control elements Control system Overall assessment No safet...

Страница 39: ...service department FAULT POTENTIAL CAUSE REMEDY The electrical height adjustment of the treatment chair does not work The electrical adjustment of backrest seat element and legrest does not work Defec...

Страница 40: ...em Angle 23 85 Fmax 90 kg Length 710 mm Width 600 mm Seatrest adjustment Electrically powered drive system Angle 28 3 Fmax 240 kg Length 510 mm Width 600 mm Footrest adjustment Electrically powered dr...

Страница 41: ...ssessment procedure in accordance with Annexes II and III Regulation EU 2017 745 Art 52 7 in the US FDA Class I Conformity assessment procedure in accordance with 21 CFR Part 886 510 k Exempt Degree o...

Страница 42: ...0 ECO_SN0001 following_0520 500 ECO www ufsk osys com User and Service Manual Page 42 of 50 10 Technical Drawing and Spare Parts List 10 1 Technical Drawing Design drawing of the 500 ECO head back sea...

Страница 43: ...500 ECO Manual_500 ECO_SN0001 following_0520 www ufsk osys com User and Service Manual Page 43 of 50 Design drawing of the 500 ECO chassis lifting column...

Страница 44: ...0041 1 6 Double castor D35 4000 2429 1 7 Bolt for double castor 4000 2430 1 8 Hexagon shaft 4000 2431 1 9 Shifter 4000 2433 1 11 Double step lever 4000 2465 1 12 Switch mode power supply 4000 2434 1...

Страница 45: ...t axle 4000 2452 4 3 Armrest left 4000 1002 4 4 Armrest right 4000 1003 4 6 Backrest covering white 4000 2454 4 7 Clamp handle grey 4000 1460 4 8 Pipe bend for headrest 4000 2455 4 10 Headrest plate 4...

Страница 46: ...department please be sure to state the equipment s name type and year of construction This information can be found on the equipment s nameplate The meanings of the label elements are explained in Se...

Страница 47: ...d technicans To connect the optionally available foot keypad please observe the illustrations and instructions below 1 Move the backrest element back to approx 30 switch off the device and remove the...

Страница 48: ...ectronic Equipment WEEE and must not be disposed of with household waste Waste electric and electronic equipment contains valuable recyclable materials that should be recycled Batteries must not be re...

Страница 49: ...annex to which this declaration refers is in due compliance with Conformity assessment procedure in accordance with Annexes II and II Regulation EU 2017 745 Art 52 7 Conformity assessment procedure A...

Страница 50: ...etachable incl adapter 3065 0000 I V Pole stainless steel 25 18mm onehand height adjustable 1070 1660mm 4 hooks max 2kg 84 4lbs hook glass and holder 3068 0000 Padded arm board with universal joint wi...

Отзывы: