UFi 1010 Скачать руководство пользователя страница 2

 

 

Cautions 
 

The UFI Model 1010 Pulse Transducer is a rugged 
instrument.  However, it may be damaged unless 
the following cautions are observed

 

Do not expose Model 1010 to temperatures 
above 50

o

 C (122

 o

 F). 

 

Do not use an ohmmeter to measure 
resistance of Model 1010. 

 

Do not apply DC voltages to Model 1010.  
These can impose large stresses that can 
bend diaphragm or fracture piezo element. 

 

Model 1010 can produce output pulses of 
several volts during handling and placement.  
If connected equipment cannot handle such 
inputs, Model 1010 should be disconnected 
from downstream equipment until it has been 
attached to subject.

 

 

Connection considerations 

 

 

Model 1010 requires that downstream signal 
conditioners have an input impedance of at 
least l00 K

 and a frequency response 

extending below 1 Hz.  A charge amplifier  
connected between Model 1010 and signal 
conditioner will extend low response. 

 

 

 

Signals of just 4-5 mV saturate some 
amplifiers.  If Model 1010 does this, a simple 
resistive voltage divider (see below) should 
be inserted between the transducer and 
instrument input.  Circuit should be shielded 
to prevent pickup of power line noise. 

 

Model 1010

1M

SHIELD

Output to signal conditioner

100K to div ide input by  10

BLACK LEAD

10K to div ide input by  100

.

 

 
Applications for the Model 1010 

 
To use as a pulse transducer: 

 

attach Model 1010 to tip of any finger with 
VELCRO® strap; 

 

lightly rest finger on transducer surface with 
edge of  finger slightly overlapping edge of 
transducer as shown in drawing below; 

 

wrap VELCRO® strap around finger. Model 
1010 should fit snugly but not restrict  
circulation.  If finger tip turns white or feels 
cold, then unit is too tight. 

Отзывы: