Fuse Replacement

The fuse is in series with the “OHM” input jack. If this fuse is open the
OHM circuit will not work. When replacing the fuse be sure to replace it
with a fuse of the same current rating and internal resistance. The use
of a fuse with a different internal resistance may cause the accuracy of
thee “

OHMS

” scale to be off. Remove the screws in the rear of the case

for access to the fuse. Fuse should be 0.3 Amp / 0.9 OHM resistance. 

Mechanical Zero Adjust

The pointer is set to register 0 at the left edge of the scale when there 
is no input to the MCP9A and it is laying face up on a flat surface. If 
the pointer does not register 0, it may be reset to that position by 
carefully adjusting the black screw in the meter face, just below the

LOCK

” switch.

S p e c i f i c a t i o n s

Optional Accessories

Optional

Battery 1.5V, size “AA”  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AB1
Fuse 0.3 Amp  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .AF45
Test leads ACV, twist lock (pr)  . . . . . . . . . . . . . . . . . . . . . . . . . . .ATL30
Test leads OHMS, twist lock (ea) . . . . . . . . . . . . . . . . . . . . . . . . .ATL35
Alligator clip adapters, insulated (pr)  . . . . . . . . . . . . . . . . . . . . .AAC
Plug-in line splitter  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ASL1

MCP9A-MAN    

P. 3

Ranges

AC Amps

0 - 6, 15, 40, 100, 300A

AC Volts

0 - 150, 300, 600V

OHMS

1K

Ω

(20

Ω

center scale)

Accuracy

AC Amps

±3% of full scale

AC Volts

±3% of full scale

OHMS

±3% of scale length

General

Jaw opening

1 - 1/4”

Battery

One 1.5 V, size “AA” (NEDA #15D) battery supplied

Fuse

0.3 Amp 0.9 OHM

Size

8 - 1/8” x 3 - 1/8” x 1 - 1/2” (206 mm x 79 mm x 38 mm)

Содержание MCP9A

Страница 1: ...MCP9A INSTRUCTION MANUAL ...

Страница 2: ...the victim s apparent recovery Electrical shock can cause an unstable heart rhythm that may need medical attention Higher voltages and currents require greater awareness of physical safety hazards Before connecting the test leads turn off power to the circuit under test set the meter to the desired function and range connect the test leads to the meter first then to the circuit under test Reapply ...

Страница 3: ...s Measuring AC Current When measuring current always be sure the SEL ECTOR switch is set to the appropriate AMP range When in doubt always use the highest current range first The switch can always be reset to a lower range for a more precise reading AC current measurements are made on the black AMP scales Clamp the jaw the MCP9A around a single circuit wire NOTE Clamp only one wire at a time for m...

Страница 4: ...MCP9A and it is laying face up on a flat surface If the pointer does not register 0 it may be reset to that position by carefully adjusting the black screw in the meter face just below the LOCK switch Specifications Optional Accessories Optional Battery 1 5V size AA AB1 Fuse 0 3 Amp AF45 Test leads ACV twist lock pr ATL30 Test leads OHMS twist lock ea ATL35 Alligator clip adapters insulated pr AAC...

Страница 5: ... implied warranties including but not limited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or consequential damages expenses or economic loss or for any claim or claims for such damage expenses or economic loss A purchase receipt or other proof of origi...

Отзывы: