Changing Scales

To change scales between degrees Fahrenheit 
and degrees Celsius, press the button on the panel, 
(marked °C / °F), while the display is active. Even if the 
trigger is released and the display is in its 60-second 
automatic hold, you can convert the reading between 
scales. Each time you press the scale button, the 
60-second hold will reset. The INF165 will default to 
the scale last used the next time it is turned on.

Using the Laser Sight

The INF165 is classified as a “Laser Product”, and is 

 

regulated by the FDA. The laser and backlight are 
always on when the trigger is pulled.

           

CAUTION!

Using controls, making adjustments, or performing  

procedures in any manner other than that specified 

herein may result in hazardous laser radiation exposure.

LCD Error Messages

The thermometer incorporates visual diagnostic  
messages as follows:

 : 

“Er2”

 is displayed when the  

 

  thermometer is exposed to rapid  

 

  changes in the ambient temperature.

 : 

“Er3”

 is displayed when the ambient  

 

  temperature exceeds the operation  

 

  range of 32˚ to 122˚F (0˚ to 50˚C). The 

 

 

  thermometer should be allowed  

 

  plenty of time (minimum 30 minutes) 

 

 

  to stabilize to the working/room  

 

 temperature.

 :  For all other error messages it is  
 

  necessary to reset the thermometer.

To reset the thermometer, turn the instrument off, remove 
the battery and wait for a minimum of one minute. 
Reinsert the battery and turn it on. If the error message 
remains please contact UEi service department for  
further assistance.

Battery Indicators

The thermometer incorporates visual low battery  
indications.

 : 

“Battery OK”

 measurements  

 

  are possible

 : 

“Battery Low”

 battery needs to be  

 

  replaced, measurements are  

 

  still possible

 : 

“Battery Exhausted”

 measurements  

 

  are not possible

Maintenance

Periodic Service

           

 WARNING!

Repair and service of this instrument is to be performed by  
qualified personnel only.

Cleaning

Periodically clean your instrument’s case using a damp cloth. 

DO NOT use abrasive, flammable liquids, cleaning solvents, 
or strong detergents as they may damage the finish, impair 
safety, or affect the reliability of the structural components.

Battery Replacement

Always use a fresh replacement battery of the specified size 
and type. Immediately remove the old or weak battery from 
the meter and dispose of it in accordance with your local  
disposal regulations. Old or defective batteries can leak  
chemicals that corrode electronic circuits.

To replace the battery:

     1.  Slide the battery cover straight down then away from  
    the INF165.

     2.  Replace with fresh AAA batteries.

     3.  Replace cover.

Lens Care

The sensor lens is the most delicate part of the thermom-
eter. The lens should be kept clean at all times. Care should 
be taken when cleaning the lens using only a soft cloth or 
cotton swab with water or medical alcohol. Allowing the 
lens to fully dry before using the thermometer. Do not sub-
merge any part of the thermometer in liquids.

            

WARNING!

Under 

NO

 circumstance should you expose batteries to 

extreme heat or fire as they may explode and cause injury.

NOTE:

 This instrument contains no user serviceable parts. If 

service becomes necessary, call UEi and ask for the service 
department. See the warranty section of this manual for 
additional details.

Specifications

INF165

Infrared Thermometer

Limited Warranty

The INF165 is warranted to be free from defects in materials and 
workmanship for a period of three years from the date of purchase. 
If within the warranty period your instrument should become 
inoperative from such defects, the unit will be repaired or replaced 
at UEi’s option. This warranty covers normal use and does not cover 
damage which occurs in shipment or failure which results from 
alteration, tampering, accident, misuse, abuse, neglect or improper 
maintenance. Batteries and consequential damage resulting from 
failed batteries are not covered by warranty. 

Any implied warranties, including but not limited to implied warranties 
of merchantability and fitness for a particular purpose, are limited to 
the express warranty. UEi shall not be liable for loss of use of the 
instrument or other incidental or consequential damages, expenses, or 
economic loss, or for any claim or claims for such damage, expenses or 
economic loss. A purchase receipt or other proof of original purchase 
date will be required before warranty repairs will be rendered. 
Instruments out of warranty will be repaired (when repairable) for a 
service charge. Return the unit postage paid and insured to:

1-800-547-5740 • FAX: (503) 643-6322

Service: (800) 308-7709

www.ueitest.com • Email: [email protected]

This  warranty  gives  you  specific  legal  rights.  You  may 

also have other rights which vary from state to state.

EMC/RFI

Readings may be affected if the unit is operated within 
radio frequency electromagnetic field strength of 

approximately 3 volts per meter, but the performance 

 

of the instrument will not be permanently affected.

IR Thermometers

Infrared Thermometers measure the amount of thermal 
radiation emitted from an objects surface. To take a 
measurement, the object must be within the field of  
view of the infrared thermometer, or the reading may 
include objects surrounding your target. This target size  
is determined in the distance to spot ratio stated on the 
instruments specifications. At a ratio of 10:1, a distance 
from your target of ten feet will give you a circle with 
one foot diameter. At 20:1 ratio gives the same one foot 
target now at 20 feet, so you can see that higher is bet-
ter, but typically more expensive due to the cost of the 
optics involved.

The other common question is regarding emissivity. In 
simple terms emissivity is a percentage of energy  
emitted from a surface compared to the energy emitted 
from a black body source. If a surface emits one half 
(1/2 or 0.5) the amount of energy at a given temperature  
and wavelength as a black body, it is said to have an  
emissivity of 0.5.

Surfaces closest to the balck body level would be flat 
black, and those farthest from it would be mirror or 

chrome surfaces. With a set emissivity level some 

 

surfaces may measure lower than actual, because their 
surface is emitting less thermal radiation at a given  
temperature.  Although you may have a fixed emissivity 
of 0.95, most items measured will provide a reasonably 
accurate result.  Having a fully adjustable emissivity will 
give you the ability to fine tune your instrument to a  
specific application.

INF165-MAN  2/06       

Copyright © 2011 UEi

PLEASE

RECYCLE

D:S=1

2:1

6.67@80”

3@36”

2@24”

INF1

65

Infrared 

Thermomete

r

Measurement range 

-36.4˚ to 689˚F (-38˚ to 365˚C)

Operating range 

32˚ to 122˚F (0˚ to +50˚C)

Accuracy (Tobj= 0~365˚C) 

+/-2.5% of reading or (2.5˚C)  

 

4.5˚F whichever is greater

(Tobj -36.4~ 0˚C)  

± (2.5˚C(4.5˚F)

 

+ 0.05˚/ per degree)

Emissivity 

0.70, 0.95

Resolution (-9.9˚ - 199.9˚F) 

0.2˚F/0.2˚C

Response Time (90%) 

1 second

Distance Spot Ratio 

12:1

Battery life 

Standard “AAA” (x2)1.5V,  

 

14 hours continuous

Dimensions 

6.9” X 1.5” X 2.8”

Weight 

6.3 oz including 

 batteries
 

2 “AAA” 1.5 batteries

“AAA” (x2)