WARNING!

Do Not expose batteries to high temperatures of flame.

Service

This instrument contains no user serviceable parts. If service becomes
necessary, call UEi and ask for the service department. See the warranty
section of this manual for additional details.

Qualified maintenance or service facilities may request 
procedures from UEi for maintaining the INF151 laser at 
FDA regulated, factory specification. Contact UEi’s service
department at the address, phone number or web-site listed 
on the back cover.

Troubleshooting

Specification Table

Malfunction

Instrument does not
turn on

Data flashes on screen

Laser comes on but no
data appears

Temperature drifts 
from known value
in a controlled
environment

Dashes appear during
measurement

Check 

Battery check

Time delay

Battery clip

Battery voltage

Battery voltage

Object emissivity

Moisture, 
corrosion or debris
on sensor

Excessive EMF

Extreme 
temperature

Corrective Action

Replace low battery

Allow 1 second for data
to appear on-screen

Ensure clip grips 
battery posts tightly

Replace low battery

Replace low battery

Prepare surface for
infrared measurement
(e.g. - tape)

Clean sensor and allow
to air dry

Move thermometer away
from interfering field

Measure surrounding
area to see if the target
exceeds the limit

Using the Laser Sight

The INF151 is classified as a "Laser Product", and is regulated by 
the FDA. 

CAUTION!

Using controls, making adjustments, or performing procedures 
in any manner other than that specified herein may result in 
hazardous radiation exposure.

To select or deselect the laser-sighting feature, press the button 
in the middle (marked "LASER") while the trigger is pulled. When the
laser mode is active, a "laser splash" will be displayed in the upper 
right portion of the LCD screen. The laser will activate on subsequent
measurements until it is deselected using the middle button. Use of 
the laser reduces battery life and is recommended for use on an 
"as-needed" basis only. The laser will remain on for approximately 
1/2 second after the trigger is released.

Maintenance

Cleaning

Prevent contact with excessive dust, dirt and liquid contamination.
Foreign substances can interfere with trigger operation and infrared 
lens performance.

When the housing becomes soiled, use a mild detergent and a damp
cloth to clean the surface. Use caution to ensure no water or soap is
allowed inside the unit or on the infrared lens.

If the infrared lens becomes contaminated try removing any debris 
with low-pressure compressed air, such as that used on computer 
keyboards. If the contaminant can not be removed with air, use a
residue free glass cleaning solution on the end of a soft cotton swab.
The swab should be slightly damp and very light pressure should be
applied to the lens.

Battery Replacement

To replace the battery:

1.  Slide the battery cover straight down then away from 

the INF151.

2.  Remove the battery clip from the old battery - a screwdriver 

may help in removing clips from battery posts.

3.  Replace with a fresh 9-volt battery.

4.  Replace cover.

Always dispose of expired batteries in a manner consistent with
local land use regulations

Range

-40 to 700˚F

(-20 to 370˚C)

Resolution

0.1 -40 to 230˚F 

(-40 to 110˚C)

1˚ 231 to 788˚F
(111˚ to 420˚C)

Accuracy

-40 ±2% or 2˚F to 382˚F

whichever greater

393 to788˚F ±3% or 3˚F

whichever greater

±1˚F/C

<500 miliseconds

41 to 104˚F (5 to 40˚C)

1"

0.85, 0.90, 0.95

8:1

Repeatability

Response time

Operating
Temperature

Minimum spot diame-
ter

Emissivity

Distance to spot ra t i o

INF151-MAN    

P. 3