DMG100 User’s Guide

3

LCD Display

Item

Function

A

Main Numeric Display

B

Alternate Numeric Display

C

“Analog” Vacuum Level Bar Graph
Backlight Indicator.  Flashing: Temporary, Solid: Always On
Sound Indicator
Battery Level Indicator
Oil Sensor Indicator
Calibration Mode Indicator
Indicates that Temperature is Displayed on the Alternate Numeric Display (during 

calibration)
Indicates degrees Celsius on the Alternate Numeric Display (during calibration) 
Pascal Units Indicator
Millibar Units Indicator
Millitorr Units Indicator
Millimeters of Mercury Units Indicator
Micron Units Indicator

DMG100 User’s Guide

8

Calibration Test

The DMG100 should rarely require recalibration, though it may be necessary to know that your gauge 

is calibrated properly for full accuracy.  The Calibration Test mode assures you that the DMG100 is 

calibrated to factory specifications. Test the calibration as follows:

1.

Turn the DMG100 power off. 

2.

Important:

 Expose the DMG100 to atmospheric pressure.

3.

Press and hold (do not release) the 

 key for about 5 seconds.

4.

The display will show 

  if the instrument is calibrated properly.

5.

The display will show 

  if the instrument requires calibration. Please see the 

Calibration section below.

Calibration

If the Calibration Test indicates recalibration is required, the gauge may be recalibrated.  Unlike any 

other vacuum gauge, the DMG100 can be easily recalibrated to factory specifications without any 

special equipment, with the following procedure:

1.

For best results, clean the sensor with alcohol prior to calibration.  Ensure the sensor is 

completely dry before proceeding.

2.

Turn the DMG100 power off.  If necessary, install a fully charged battery into the gauge.

3.

Place the dust cap over the vacuum fitting.

4.

Hold

 while pressing 

.  As soon as the power turns on, release 

 and 

 and then 

press

 quickly at least three times.  The 

CAL

indicator should illuminate, and 

should show on the Main Numeric Display. If not, turn the power off again and repeat.

5.

Place the DMG100 in a clear Ziploc (resealable zipper storage) bag, press out any extra air, and 

seal.

6.

Place the bagged DMG100 into a freezer with a temperature of less than –5

°

C (23

°

F).

7.

Allow the DMG100 to cool to below -2

°

C (28.4

°

F). At this point, the alarm will sound and the 

display will change to 

.

8.

Remove from freezer and press any key to silence the alarm.

9.

Place the DMG100 undisturbed in an area with a room temperature of at least 23

°

C (73.4

°

F) but 

no greater than 30

°

C (86

°

F).

10. Allow the DMG100 to warm to 20

°

C (68

°

F).  At this point, the alarm will sound, and 

will show on the display.

11. Press any key to silence the alarm.  The DMG100 is now calibrated to factory specifications.

Note:

For accurate calibration, it is necessary to allow the DMG100 to warm slowly.  Attempting to 

accelerate the warming by using a heat source will not provide satisfactory results.  During the 

cooling/warming process, the temperature will be indicated on the Alternate Numeric Display in 

degrees Celsius.  The calibration process may be canceled at any time by turning off the DMG100. 

The previous calibration will be unchanged.

www.

.com

[email protected]

1.800.561.8187

Содержание DMG100

Страница 1: ...es not ch results from t or improper lting for failed plied warranties are limited to of use of the s expenses or age expenses or iginal purchase ce charge UEi or visit us o 7008 6322 om ts ate to state 643 6322 04 278 8299 DMG100 Digital Micron Gauge www com information itm com 1 800 561 8187 ...

Страница 2: ... uum port with your hand turn the BluVac back on ean the Vacuum Sensor nd Restart Calibration art Calibration Allow the Gauge to Warm Slowly start Calibration Leave he Gauge Undisturbed uring Warming Phase 12 C While operating below every second to one reading mperatures below 0 C 32 F DMG100 User s Guide 2 Parts Diagram www com information itm com 1 800 561 8187 ...

Страница 3: ...actory specifica 1 Turn the DMG100 powe 2 Important Expose the 3 Press and hold do not r 4 The display will show 5 The display will show Calibration section below Calibration If the Calibration Test indicate other vacuum gauge the DM special equipment with the fo 1 For best results clean th completely dry before pr 2 Turn the DMG100 powe 3 Place the dust cap over 4 Hold while pressing press quickl...

Страница 4: ...ater than 25 000 Microns 3 Select the desired units by repeatedly pressing 4 Attach the DMG100 to the system to be evacuated with a high quality vacuum hose Start the vacuum pump Read the vacuum level from the Main Display 5 Turn the power off by pressing and holding Battery Installation Replacement 1 Remove the battery cover from the rear of the DMG100 by compressing tab at the base of the batter...

Страница 5: ...eep twice every 5 minutes to remind you that it is on The DMG100 will not sleep in the Calibration Mode see the Calibration section on pg 8 Auto Power Off After 1 hour in Sleep Mode the DMG100 will automatically turn itself off to further conserve battery power Oil Sensor It is necessary to prevent oil from being drawn into the Vacuum Sensor For HVAC R service always recover refrigerant prior to a...

Страница 6: ... to the table below Vacuum Range Resolution Pascals Pa 0 3 333 1 0 1 Pa Millibar mBar 0 33 331 0 001 mBar Millitorr mTorr 0 25 000 1 mTorr mmHg mmHg 0 25 000 0 001 mmHg Microns Microns 0 25 000 1 Micron Analog Vacuum Level Bar Graph The DMG100 also displays a Vacuum Level Bar Graph allowing for a quick visual determination of the vacuum level achieved Each bar corresponds to a vacuum pressure rang...

Страница 7: ... wet rag Mild detergent is acceptable but use no solvents Take care not to expose the vacuum sensor to oil If the Oil Sensor see pg 5 indicates a contaminated sensor follow the Sensor Cleaning Procedure below Cleaning the Vacuum Sensor If the vacuum sensor becomes contaminated with oil as indicated by the Oil Sensor carefully follow this procedure 1 Turn the DMG100 power off 2 Shake the gauge to r...

Страница 8: ...4 Hold while pressing As soon as the power turns on release and and then press quickly at least three times The CAL indicator should illuminate and should show on the Main Numeric Display If not turn the power off again and repeat 5 Place the DMG100 in a clear Ziploc resealable zipper storage bag press out any extra air and seal 6 Place the bagged DMG100 into a freezer with a temperature of less t...

Страница 9: ...olution Sensor Contaminated Clean the Vacuum Sensor Normal Operation Ambient Temperature too Low Turn the BluVac off warm the vacuum port with your hand and turn the BluVac back on Calibration Sensor Contaminated Clean the Vacuum Sensor and Restart Calibration Gauged Warmed too Quickly Restart Calibration Allow the Gauge to Warm Slowly Calibration Gauge Disturbed During Calibration Restart Calibra...

Отзывы: