background image

When measuring resistance, be sure you have good contact between
the test leads and the circuit. Dirt, oil, solder flux, or other foreign 
matter alters the reading value.

To measure resistance, follow these steps:

1. Insert the test leads into the meter, and turn off the power to the 

circuit under test. Voltage, from any source, across the circuit will 
cause an erroneous reading.

2.  Set the rotary dial to the “

Re s i s t a n ce / C o n t i n u i t y

” function, then  

turn on th power to the meter. The meter defaults to the resistance 
measurement mode.

3.  Touch the test probes to the test points, and read the display. 

The meter beeps as it seeks the correct range to measure the 
circuit resistance.

Note:

During continuity or resistance checks, polarity does not matter.

It may test circuits to use the manual range mode to test circuits 
that are susceptible to damage at low voltages/currents. See 
the table to determine which range is best suited to your circuit.

Continuity Test Mode

Use this mode to make quick checks for continuity of electrical circuits,
such as wiring, speaker cables, connections, switches or relays. In the
continuity mode, an audible tone sounds when the value measured is
approximately 50

or less.

To test for continuity, follow these steps:

1. Set the rotary dial to the “

Ohms/Continuity

” function.

2.  Press the “

Ω

/

“  push-button to bring the “       “ symbol up 

on the top of the display, indicating that the continuity mode has 
been selected . The range is preset to the 400

Ω

s cale and can not 

be changed.

3.  Place one probe to each side of the circuit to be tested. If 

approximately 50

Ω

or less resistance is in the circuit, the meter 

sounds a continuous tone.

Diode Test Mode

The diode test function allows you to check diodes, transistors and other
semi-conductor devices for open, shorts and normal operation.

CAUTION!

Never connect the test leads to a source of voltage when the diode test
function is selected.

Diodes and other P/N junction devices, easily allow current flow in one
direction and prevent current flow in the other direction. When a diode
is forward biased, it allows current to flow. When it’s reverse biased it
prevents current flow.

Your meter is designed to apply enough voltage in the forward biased
direction to allow current flow. All diodes use up or “drop” a small
amount of the supplied voltage when they are forward biased, and drop
nearly all of the supplied voltage when they are reverse biased. The
voltage drop of the diode under test is displayed on the meter when it is
forward biased. This is normally around 0.4 V for germanium diodes
and 0.6 V for silicon diodes. When the diode is reverse biased, the
meter should indicate “.

OFL

”.

Determine the condition of semiconductor devices as follows:

1. Select the diode test function on your rotary dial.

2.  Connect the red test lead to the anode side of the diode, and the 

black to the other. A black band is normally printed around the 
end of the side of a diode.

3.  Note the displayed value.

4.  Reverse the red and black test leads and again note the 

displayed value.

5.  If the digital reading in the first (forward biased) direction 

indicates some measurable value, and the reading in the reversed 
direction shows an overload “.

OFL

”, the diode is good.

6.  If the displayed value is low or measurable in both directions, the 

diode is probably shorted.

7. If the display indicated an overload “.

OFL

” in both directions, the 

diode is probably open. Some diodes require a higher biasing 
voltage than this meter supplies. See UEi’s catalog to purchase an 
economical, high-power diode test lead adapter set if necessary.

Measuring Current (amps)

CAUTION!

The current functions are protected by 250 volt rated fuses. To avoid
damage to the instrument, do not measure current sources with open
circuit voltages greater than 250 volts DC or AC.

Note:

When taking current measurements, this meter must be 

connected in series with the circuit (or circuit element) under test. 
Never connect the test leads across a voltage source while the rotary
dial is set to amps. This can cause damage to the circuit under test 
or this meter.

The low amperage measurement functions share the test probe port
with the volt and ohm measurement.

DO NOT

turn the rotary dial to the 

mA

or 

µA

position until the circuit is

prepared for current measurements as described below.

To measure current, you must break the circuit under test and make 
the meter part of the circuit. Two connection points are created when 
a circuit is broken. On one side is the power source and the other is 
the load. Connect the two leads in the circuit at the points created by
this break.

When measuring alternating current, any waveform other than a 
sinusoidal wave will have inherent measurement error.

DM400-MAN    

P. 5

Содержание DM400

Страница 1: ...1 800 547 5740 Fax 503 643 6322 www ueitest com email info ueitest com DM400 INSTRUCTION MANUAL...

Страница 2: ...behind the finger guards molded into the probes Use a current clamp adapter when measuring current that may exceed 10 amps See the accessories in UEi s full line catalog Do not open the meter to repla...

Страница 3: ...layed when the button is pressed Each subsequent press of this button changes the range as indicated by the annunciators and the decimal point placement An audible beep is heard whenever the range cha...

Страница 4: ...d in the absence of a terminal input the range is automatically set to the following Note The 400 mV DC or AC range must be selected manually The meter will not seek this range in the autorange mode F...

Страница 5: ...voltage from phase to phase is over 831 V AC This exceeds the maximum rating of this meter and may result in personal injury or instrument damage Measuring Resistance CAUTION Turn off the power and di...

Страница 6: ...p or drop a small amount of the supplied voltage when they are forward biased and drop nearly all of the supplied voltage when they are reverse biased The voltage drop of the diode under test is displ...

Страница 7: ...could result in physical degradation of the meter This could alter the protection from electrical shock and personal injury this meter provides to the operator Perform only those maintenance tasks tha...

Страница 8: ...tinuous use for five minutes disconnect the current input from the 10 ampere mode for least five additional minutes before making a second measurement Maximum Inputs DM400 MAN P 7 Function DC Volts AC...

Страница 9: ...idity Temperature Coefficient Battery Type Battery Life alkaline Size H x W x D Meter only with Holster Weight Meter only with Holster Vibration and Shock Pollution Degree IP Code Safety Standards Cou...

Страница 10: ...imited to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental o...

Отзывы: