background image

Audible alarm

An audible alarm will sound if 35 PPM CO (or great-
er) is detected. The alarm frequency will increase 
with higher concentrations. Check the LCD display 
for actual level. While the alarm is sounding the unit 
can not be turned off.

Back-Light

To activate the back-light press the 

ON/OFF

 key 

quickly once. Do not hold down the key or the unit 
will switch off.To turn off the back-light press the 

ON/OFF

 key quickly again. 

Note:

 Using the back-light decreases battery life.

LOW BATTERY INDICATION

Low battery indicator

If the battery voltage falls below a pre-set level the  
symbol 

“BAT”

 appears in the display. Do not use the 

instrument when this symbol is present. Replace the  
battery immediately. 

Troubleshooting

If the display fails to stabilize at around zero within a  
few minutes this may indicate an expired sensor or a  
sensor that has been exposed to excessive levels of CO.  
If the symptom persists seek advice from an approved 
service center.

Always check for correct operation before each use. Use  
a known source of CO (e.g. a smoking match stick or a 
recently extinguished flame) taking care to act in a safe 
manner and not to expose the CO71A sensor to extreme 
temperatures or CO concentration. If the unit does not 
change as expected, return the unit to UEi or an approved 
service center. DO NOT test using automotive exhaust as 
the level may exceed the instrument range. This will poison 
the sensor or decrease sensor life span.

If during use, you detect CO leaks, be sure you know the 
appropriate actions to take to protect yourself and others. 
Before switching off the instrument always expose it to 
fresh air until the reading falls below 5 PPM. Note that the 
readings fall much more slowly than they rise so you may 
need to wait three minutes or longer.

Maintenance
Calibration

The CO71A must be calibrated annually at a certified 
sevice location to maintain accuracy. Electrochemical 
sensors are subject to long term drift that is corrected 
during calibration. Refer to troubleshooting section of this 
manual to check responce to CO. THIS DOES NOT verify 
accuracy.

Specifications

No user serviceable parts. Re-calibrate every 12 
months. Check for correct operation as described in 
this manual before every use.

Concentration
of CO in Air
9  ppm

35 ppm

200 ppm

400 ppm

800 ppm

1,600 ppm

3,200 ppm

6,400 ppm

12,800 ppm

Inhalation Times and Toxic 

Symptoms*

The maximum allowable 

concentration for short term 

exposure in a living area 

according to ASHRAE.

The maximum allowable 

concentration for 

continuous exposure in any 

eight hour period industry 

according to MIOSHA.

Maximum concentration 

allowable in any 15 minute 

period according to OSHA. 

Slight headaches, fatigue, 

dizziness, nausea after 2 to 

3 hours.

Frontal headaches within 1 

to 2 hours, life threatening 

after 3 hours. Maximum 

allowable limit in flue gas 

according to EPA and AGA.

Dizziness, nausea and 

convulsions within 45 

minutes. Unconsciousness 

within 2 hours. Death within 

2 to 3 hours.

Headache, dizziness and 

nausea within 20 minutes. 

Death within 1 hour.

Headache, dizziness and 

nausea within 5 to 10 

minutes. Death within 30 

minutes.

Headache, dizziness and 

nausea within 1 to 3 

minutes. Death within 10 to 

15 minutes.

Death within 1 to 3 minutes.

*The effects can vary depending on size, age, sex and health. UEi 
is not liable for the accuracy of this information.

  (Table 1)

039

BAT

Range: 

0-999 ppm

Resolution: 

1 ppm

Accuracy: 

±(3% of r 1 ppm)

Audible Alarm: 

Preset at 35 ppm

Visual Indication: 

LCD and LED

Battery: 

9V alkaline (NEDA 1604)

Ambient

Temperature: 

32 - 104˚F (0 - 40˚C)

Warranty: 

3 year

Отзывы: