C50/C75

INSTRUCTION MANUAL

99 Washington Street 
Melrose, MA 02176 

Phone

 781-665-

1400

Toll Free 1-800-517-8431

Visit us at www.T

estEquipmentDepot.com

Содержание C75

Страница 1: ...C50 C75 INSTRUCTION MANUAL 99 Washington Street Melrose MA 02176 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com...

Страница 2: ...N and OFF Hold to scroll through top line display also selectable through menu 6 Hold Store Freezes reading on display Entire display flashes Press and hold for 2 seconds to store data in memory 7 Rot...

Страница 3: ...s the ratio of measured CO divided by calculated CO2 It gives an indication of How good a gas sample the instrument is reading How clean the boiler is running For example A new or clean domestic boile...

Страница 4: ...he sampling point be located at least two flue diameters downstream of any bend as close to the source as possible and that the probe tip is in the center of the flue With balanced flues and other dom...

Страница 5: ...above the function keys The three keys are increases decrease and enter Set Time 1 Press ENTER 2 Use the up and down keys to select the correct time 3 Press ENTER to move to the next digit Note Time...

Страница 6: ...TORE view mode the rotary dial is used to select the saved parameter to be displayed Use the UP DOWN keys to change the test number To exit the view mode press ENTER Storing Test Results At any time d...

Страница 7: ...x 2488 2 1 Tf 4 2 Ti Simplified 9 x H2 H2O Qgr x 2425 x 1 0 001 Tnet Wet loss K3 1 0 001 x Tnet Where K3 9 x H2 H2O Qgr x 2425 Combustion Combustion Theory In its simplest form combustion is the comb...

Страница 8: ...r free In the example we would be breaking the limit and corrective action should be taken to reduce the level of CO Air free values prevent false readings being submitted e g allowing more air into t...

Страница 9: ...Slide outward Pull plug out to drain Insert a new filter Maintenance Periodic Service WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or s...

Страница 10: ...10 V AC off load Annual Re Calibration While the sensor has an expected life of more than two years in normal use it is recommended that the analyzer is re calibrated at least annually This is so that...

Страница 11: ...ted to implied warranties of merchantability and fitness for a particular purpose are limited to the express warranty UEi shall not be liable for loss of use of the instrument or other incidental or c...