• Remove the protective rubber boot

• Slide the water trap unit from the analyzer

• Remove the particle filter from its’ spigot and replace

• Slide the water trap back into position and replace protective rubber boot

• Remove the rubber plug

• Allow the water to drain out

• Re-insert the rubber plug

FRESH AIR PURGE

Each time the analyzer is turned on it will perform a 60 second air purge, this is to clear 

the gas sampling path (including probe, if connected), and self calibrate the CO

2

   

detection system. For these reasons it is very important that the analyzer be in 

outside 

fresh air 

when powered on.

  

WARNING

When in the Rotary Dial positions O2/Eff; Press the Up or Down Buttons, scroll to CO2 

Zero. On the Purge option,  short press the Edit Button, and a 60 second Purge will begin.

STATUS

FLUE 1

AUX

FLUE 2

TEMP

Combu

stion Analy

zer

C161

CO

CO

2

TE

M

P

EMPTYING & CLEANING THE IN-LINE WATER TRAP

CHANGING THE PARTICLE FILTER

QUICK START

Turn on the analyzer by pressing the On/Off Button for 2 seconds until the unit activates. As described the analyzer will perform a 60 

second turn on purge. Once completed simply select the reading desired by the position of the rotary dial.

MEASURING FLUE GASES

After the initial countdown is finished and the analyzer is properly setup, put its’ flue probe in the appliance’s sampling point. The tip of 

the probe should be at the center of the flue. Use the flue probes depth stop cone to set the position.

With balanced flues, make sure the probe is positioned into the flue so no air can “back flush” into the probe.

NOTE: 

Ensure the flue probe handle does not get hot!

20”

Boiler

Flue

Содержание C160

Страница 1: ...AL ENGLISH Flue Gas Combustion Analyzer Combustion Analyzer C160 CO C O 2 T E M P Status v3 O2 Eff CO CO2 5 Commonwealth Ave Woburn MA 01801 Phone 781 665 1400 Toll Free 1 800 517 8431 Visit us at www TestEquipmentDepot com ...

Страница 2: ...r Icon Layout 10 Status Bar Menu Options 10 Status Bar Options Table 10 Menu 11 Using the Menu 11 Menu Items 11 Stored Logs Memory 12 Menu Options 12 Viewing Stored Logs 13 Log View Menu Options 13 Navigating Stored Logs 13 Log Navigation Menu Options 14 Printouts 15 Specifications 16 Certification 16 Where to Test 17 18 What Results are Generally Acceptable 18 What Results Are Generally Acceptabl...

Страница 3: ...s who are familiar with the hazards of their trade Observe all recommended safety procedures that include proper lockout utilization and use of personal protective equipment that includes safety glasses gloves and flame resistant clothing Users of portable gas detectors are recommended to conduct a bump check before relying on the unit to verify an atmosphere is free from hazard A bump test is a m...

Страница 4: ...ion if Wireless Module is fitted K Down Button Short press to navigate DOWN L ENTER Button Short press select current option displayed Long press to activate some menu items M Rotary Dial N Particle Filter inside water trap O Water Trap P LED Full Water Trap Indicator Q Serial Number under Protective Boot R Sensors Fitted under Protective Boot Indicates Sensors fitted in unit S Battery Compartment...

Страница 5: ...T E M P Status v3 O2 Eff CO CO2 Serial No Sensors Fitted CO O2 T1 B C D E F G H I J K L M W V X P O N S U T R Q Flue Probe Temperature Plug Plugs into T1 Narrow Pin MUST be on the Right hand side Flue Probe Gas Inlet Plug Black Connector plugs into Flue Gas Inlet ...

Страница 6: ... system Clean particle filter Water trap and probe line are empty of water Power on and zero All hose and thermocouple connections are properly secured Flue gas probe is sampling ambient FRESH air Water trap is fitted correctly to the instrument Flue temperature plug is connected Ensure water trap plug is inserted correctly WARNING Turning the pump off while the probe is in the flue will leave tox...

Страница 7: ...Zero On the Purge option short press the Edit Button and a 60 second Purge will begin STATUS F L U E 1 AUX FLUE 2 T E M P P EMPTYING CLEANING THE IN LINE WATER TRAP CHANGING THE PARTICLE FILTER QUICK START Turn on the analyzer by pressing the On Off Button for 2 seconds until the unit activates As described the analyzer will perform a 60 second turn on purge Once completed simply select the readin...

Страница 8: ... Francais EFF Efficiency select between Efgn or Efgc except for Heavy Oil and Pellets O2 Ref Selectable parameter select between 0 0 to 20 0 Logs if logs are saved they will be displayed Code for Authorized Service Personnel only Status Screen Ti Ambient temperature CAL days until next calibration due O2 Eff Screen O2 displayed in Efficiency Efgn Net efficiency Efgc Gross efficiency T1 displayed i...

Страница 9: ...no activity Navigation through the various start options and menu system is via the 3 button dedicated Up Down Enter Buttons Button presses are categorized short press and long press in the same manner as many modern consumer electronics Status Bar The Status Bar provides the user with a concise indication of the instrument status it incorporates an intelligent system that understands what the use...

Страница 10: ...u Options The Status Bar will offer the user Menu items based on the Selector Dial setting Status Bar Options Table Test Screen Dial Digital Position Test Screen Status Info Available STATUS O2 Eff CO CO2 TIME 4 4 4 FLUE SELECTED 4 4 4 DATE 4 4 4 CO2 FLUE 4 4 Date Option Date Displayed as Pump Pump Sending Data Battery Pump on Pump paused Indicates data is being transmitted to either the App or th...

Страница 11: ...roll fields to change Press and buttons to change field contents Press to enter content selected Menu Items MENU Sub Menu OPTIONS COMMENTS Time HH MM SS 24 Hour Format e g 7AM 0 00 00 7PM 19 00 00 Date DD MM YY Header LINE 1 Customer info displays on print outs LINE 2 Customer info displays on print outs EXIT IR Print KMIRP Select to print to KMIRP IRP 2 Select to print to IRP 2 Temperature Scale ...

Страница 12: ...ed for Logs of any type To view the current memory usage status simply navigate to menu LOGS press the button to select the Log desired Number of stored Combustion Logs Total Memory Used Menu Options Menu Options Combustion Analyzer C160 CO C O 2 T E M P Status v3 O2 Eff CO CO2 View stored Logs this will enter the secondary menu to select the type of log to view will only appear if data available ...

Страница 13: ...nce the desired Log type to view has been selected the first available stored Log will be displayed Log Number Log time and date this line will alternate between time and date of Log Log readings information is specific to Log type Navigation Menu options STATUS F L U E 1 AUX FLUE 2 T E M P Combustion Analyzer C161 CO C O 2 T E M P View stored Combustion Logs Option will appear if Logs are availab...

Страница 14: ... Log Navigate to the next available Log of the selected type Option will only appear if there is more than one Log Navigate to the previously selected Log of the selected type Option will only appear if there is more than one Log Exit back to the previous menu ...

Страница 15: ... Printing To print simply press and release the PRINT BUTTON Printing can be aborted by pressing the PRINT BUTTON again UEi Infrared Printer Make sure the printer is switched on ready to accept data and its infrared receiver is in line with the analyzer s infrared emitter on top of the analyzer Combustion UEi C160 SW20157 v1 1 0 11 20 76 1 0 0009 83 9 73 9 c c T1 Ti NETT EFFgc LOSS XAIR 08 19 21 0...

Страница 16: ...0379 Parts 1 3 in accordance to 1st German Feder al Emission Control Ordinance Bim5chV Operating Conditions Temperatures Humidity 32 to 113 F 0 to 45 C 15 to 90 RH non condensing Power Supply Rechargeable batteries USB Charging Physical Characteristics Weight Dimensions 1 5 lb 635g H 8 54 x W 4 18 x D 1 86 H 217 mm x W 106 mm x D 47mm The C160 is in conformity with the relevant Union harmonization...

Страница 17: ...ers Draft Verify proper combustion O2 CO2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Supply and Return Water temp 80 Furnace Verify proper combustion O2 CO2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit Switch Pressure Switch Verify proper operation Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils 90 Furnace Verify proper combust...

Страница 18: ...tatic Duct Pressure Temperature Rise AC side Static Pressure Drop across coils WHAT RESULTS ARE GENERALLY ACCEPTABLE What Results Are Generally Acceptable Atmospheric Gas Fixed Burners Gas Fired Burners Oil Fired burners 2 Oil Fuel Positive Overfire Gas Oil Oxygen 7 to 9 O2 3 to 6 O2 4 to 7 O2 3 to 9 O2 Stack temperature 325 F to 500 F 275 F to 500 F 325 F to 600 F Draft Water Column Inches 02 to ...

Страница 19: ...ry immediately Liquids can degrade electronic circuits Whenever practical keep the instrument away from dust and dirt that can cause premature wear Although your instrument is built to withstand the rigors of daily use it can be damaged by severe impacts Use reasonable caution when using and storing the meter COLD WEATHER PRECAUTIONS It is important you keep your flue gas analyzer in a warm and dr...

Страница 20: ...lyzer remove the protective rubber boot and fit 3 AA NiMH rechargeable batteries in the battery compartment Take great care to ensure they are fitted with the correct battery polarity Replace the battery cover and the protective rubber boot Time and Date When changing the batteries the time and date will need setting Battery Disposal Always dispose of depleted batteries using approved disposal met...

Страница 21: ...omers book their recertification they receive a prepaid UPS Ground shipping label PRODUCT REGISTRATION Register Online Registering you analyzer online is quick and easy Just log in or setup an account it only takes a couple of minutes Once logged in you can register you analyzer by providing some product information and uploading a proof of purchase When it s time to request recertifcation just lo...

Страница 22: ...easure the CO2 level and the stack temperature and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate the efficiency as adjustments are being made Combine this with a printout and you are able to provide a before and after comparison of the combustion equipment in rel...

Страница 23: ......

Страница 24: ...will be repaired or replaced at UEi s option This warranty covers normal use and does not cover damage which occurs in shipment or failure which results from alteration tampering accident misuse abuse neglect or improper maintenance Batteries and consequential damage resulting from failed batteries are not covered by warranty Any implied warranties including but not limited to implied warranties o...

Отзывы: