UEi 053533507225 Скачать руководство пользователя страница 12

12

GENERAL MAINTENANCE

PRE TEST CHECKLIST

SETTING INLET TEMPERATURE

ANALYZER CONNECTIONS

EMPTYING & CLEANING THE IN-LINE WATER TRAP

CHANGING THE PARTICLE FILTER

•  Re-certify your instrument annually to ensure it meets original performance specifications
•  Keep your instrument dry . If it gets wet, wipe dry immediately . Liquids can degrade electronic circuits
•  Whenever practical, keep the instrument away from dust and dirt that can cause premature wear
•  Although your instrument is built to withstand the rigors of daily use, it can be damaged by severe impacts . Use reasonable caution  

 

when using and storing the meter

•  Turn on and zero the analyzer with out the flue probe connected to use ambient temperature
•  Connect flue probe thermocouple to T1 during zero countdown to store probe tip temperature as inlet (ducted system)

The in-line water trap should be checked and emptied on a regular basis . Water vapor will condense in 

the probe line, which may cause the water trap to fill suddenly if the probe is moved . Care should be 

taken at all times .

Carefully remove the rubber plug from the bottom of the water-trap housing . Dispose of the condensate 

in a suitable drain, care must be taken as it could be acidic . If condensate spills onto the skin or clothing, 

clean off immediately using fresh water, seek medical advice if problems occur . Ensure plug is replaced 

before performing combustion tests . Note: O2 reading will be high if the Water Trap Plug is not in place .

This is a very important part of the analyzer and should be changed regularly . It prevents dust and dirt 

particles from entering the pump and sensors that will cause damage . The filter MUST be changed when 

it appears discolored on the inner surface .

Remove water-trap assembly from the analyzer as shown above . Remove the filter and plastic holder 

from the housing . Discard the filter element but keep the holder to fit to the new filter . Clean the inside 

of the filter housing with a suitable soft cloth . Fit the holder onto the new filter element and then insert 

into the housing . Refit the housing onto the analyzer .

•  Clean particle filter 
•  Water trap and probe line are empty of water
•  Power on and zero
•  All hose and thermocouple connections are properly secured

•  Flue gas probe is sampling ambient FRESH air
•  Water trap is fitted correctly to the instrument
•  Flue temperature plug is connected

WARNING!

Turning the pump off while the probe is in the flue will leave toxic gases inside 

the analyzer . Once data has been printed or copied it is advisable to purge the 

unit with fresh air as soon as possible . To do this remove the probe  from the flue 

and turn ON the pump . Always allow the readings to return to zero (20 .9 for O2) 

prior to shutting the unit off .  The meter will not switch off until the CO reading 

is below 20 ppm .

WARNING!

The probe will be hot from flue gases .  Remove the probe from the flue and allow 

it to cool naturally . Do not immerse the probe in water, as this will be drawn into 

the analyzer and damage the pump and sensors . Once the probe is removed from  

the flue and the readings have returned to ambient levels hold down “On/Off” 

and switch off the analyzer . The instrument will count down from 30 to switch 

off . If you pressed “On/Off“ by mistake, pressing “Send“ will return you to normal 

operation .

NOTE: Take care when inserting 

the temperature probes as the 

pins are polarized . Insert with the 

smaller pin (+) to the right . 

NOTE: Do not insert anything 

into programming port.

www.

.com

[email protected]

1.800.561.8187

Содержание 053533507225

Страница 1: ...I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time COp 28 CO2 6 2 C85 INSTRUCTION MANUAL ENGLISH Long Life EOS Analyzer www com information itm com 1 800 561 8187 ...

Страница 2: ...odic Service 13 Analyzer Re Certification 13 Cleaning 13 Other Important Factors Related To Combustion 14 Combustion Measurement Terms 14 Specifications 15 Electromagnetic Compatibility 15 Limited Warranty 16 DISPLAY SYMBOLS Excess Air Theoretical amount of air in excess of level needed to completely burn fuel Delta T Differential temperature Losses Losses calculated from oxygen and type of fuel I...

Страница 3: ...igate DOWN SEND ENTER to Select and Print Particle Filter Inside Water Trap Water Trap Protective Rubber Boot w built in Magnets Battery Charge AC Adapter Flue Gas Inlet Connection Water Trap Drain 2 Line Backlit Display On Off Button Rotary Test Selector Dial Battery Compartments in back under boot AC Power Charge Indicator Light Temperature Connections Flue Probe Temp T1 Flue Probe Temperature P...

Страница 4: ...nce all digits are set Press ENTER to complete setting the date once all digits are set C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time MENU SET TIME C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time SET TIME 08 30 15 _ C85 L O N G L I F E E O S A N A L ...

Страница 5: ...O Temp Fuel Date Time AUXPAGE1 EXIT C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time AUX PAGE2 C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time AUX EXIT AUXPAGE1 LINE 2 shows press Send to continue to set line 2 Press s UP or t DOWN select between availa...

Страница 6: ...CO Temp Fuel Date Time HEADER HEADER2 C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time _PHONE NUMBER C85 L O N G L I F E E O S A N A L Y Z E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time HEADER EXIT The existing HEADER 1 will display with the cursor underneath the first digit Press s UP or ...

Страница 7: ...E R CO2 CO TEMP Menu Status Aux Eff X Air O2 CO Air Free CO2 CO Temp Fuel Date Time DEL ALL YES PRINT SETUP Rotate dial to Menu Rotate dial to Menu Press s UP or t DOWN to REPORT Press ENTER Press s UP or t DOWN to PRINTING Press ENTER Press s UP or t DOWN to DEL ALL Press ENTER to proceed Press s UP or t DOWN to select printer Press ENTER to select Analyzer will automatically return to the menu P...

Страница 8: ... is approximately at the center of the stack 4 stack adjust cone to aprox 2 from end of probe Rotate selector to desired test Rotate selector to Status The status menu shows battery life and the number of days until suggested re certification CAL Rotate selector to Aux The auxilliary menu shows user defined parameters The display alternates between page 1 and page 2 To set up the auxilliary menu s...

Страница 9: ...Free The O2 CO Air free screen shows the calculated amount of O2 from CO2 measured and fuel selected and CO Air Free CO Air free is a calculated value for carbon monoxide as if there were no dilution from air COppm 20 9 20 9 O2 measured Rotate test selector to Fuel The fuel selector position will display the current fuel being used for calculations Press s UP or t DOWN to scroll and press ENTER to...

Страница 10: ...tion O2 CO Stack Temp Efficiency Hi Low fire Gas Pressure Send and Return Water temp 80 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Efficiency Verify Set Up Gas Pressure Test Limit Switch Pressure Switch Verify proper operation Static Duct Pressure Temperature Rise AC side Static Pressure Drop across coils 90 Furnace Verify proper combustion O2 CO Stack Temp Vent Pressure Effic...

Страница 11: ...o 9 O2 Stack Temperature 325 to 500 F Draft Water Column Inches 02 to 04wc Carbon Monoxide parts per million 100ppm Gas Fired Power Burners Oxygen 3 to 6 O2 Stack Temperature 275 to 500 F Stack Draft Water Column Inches 02 to 04wc Overfire Draft Water Column Inches 02wc Carbon Monoxide parts per million 100ppm Oil Fired Burners 2 Oil Fuel Oxygen 4 to 7 O2 Stack Temperature 325 to 600 F Stack Draft...

Страница 12: ...ter MUST be changed when it appears discolored on the inner surface Remove water trap assembly from the analyzer as shown above Remove the filter and plastic holder from the housing Discard the filter element but keep the holder to fit to the new filter Clean the inside of the filter housing with a suitable soft cloth Fit the holder onto the new filter element and then insert into the housing Refi...

Страница 13: ...stent with local disposal regulations WARNING Repair and service of this instrument is to be performed by qualified personnel only Improper repair or service could result in physical degradation of the instrument This could alter the protection from personal injury this meter provides to the operator Perform only those maintenance tasks that you are qualified to do Remove the probe from the flue a...

Страница 14: ... and the stack tempera ture and then reference a slide scale that would give you the relative efficiency number UEi s electronic combustion analyzers perform the measurements on a continuous basis and can calculate the efficiency as adjustments are being made Combine this with a printout and you are able to provide a before and after comparison of the combustion equipment in relatively little time...

Страница 15: ...s certified to be compliant The European Council Directive 89 336 EEC requires that electronic equipment does not generate electromagnetic disturbances that exceed defined levels and has an adequate level of immunity to enable it to be operated as intended Since there are many electrical products in use that pre date this Directive and may emit electromagnetic radiation in excess of the standards ...

Страница 16: ...Canadian customers we have a recertification service center in Vancouver BC which is ISO IEC 17025 2005 accredited and offers NIST certification services for select products All recertification services are warranted for one year Save time with online recertification check in by visiting www ueitest com and click the SERVICE REQUEST banner and filling out the form Our current recertification price...

Отзывы: