background image

Digital Medical Thermometer

0833

PZN 5880276

Instructions for use

0123

Flexible tip

Figure example

SAFETY INSTRUCTIONS

MEASURING THE TEMPERATURE

MEASUREMENT TYPES

In the rectum (rectal measurement)

In the mouth (oral/sublingual measurement)

Under the arm (axillary measurement)

Display

On/off button

Battery compartment cover

Carefully insert the tip 1-2 cm into the anus. Minimum measurement time: 30 seconds

Do not consume hot or cold food or drinks for at least 10 minutes before measuring body temperature. Insert 

the tip into one of the two heat pockets under the tongue, on the left- or right-hand side of the root of the 

tongue. Close the mouth and breathe calmly through the nose. Minimum measurement time: 30 seconds

The skin must be completely dry. Place the tip under the arm, so that it is fully in contact with the skin. The 

thermometer should be at right angles to the body. Place the arm across the chest, so that the tip is tightly 

covered. Minimum measurement time: 3 minutes.

To turn on the thermometer, press the on/off button. A short beep indicates that the thermometer is on. At 

the same time, an optical full segment check of the display will be carried out. Then the last temperature 

measured will be displayed for approx. 2 seconds, along with a small M (for memory). Then an internal test 

value of 37.0 °C appears and the thermometer switches to measuring mode. Now you can take a new 

measurement.

For temperatures up to 37.5 °C, the normal sound will be made (10 long, equal beeps). For temperatures 

over 37.5 °C, the fever alarm will sound (30 short beeps in groups of 3).

Read the temperature from the display. The thermometer will switch itself off automatically after approx. 10 

minutes. You can also turn it off earlier by pressing the on/off button. The temperature will be saved until 

the next one is measured.

Place the thermometer at the desired measurement point. During measurement the current temperature is 

shown on the display with a flashing „°C“ symbol. If the increase in temperature over a period of 16 sec-

onds is less than 0.1 °C, the thermometer will beep. If the temperature is being measured orally or rectally, 

this beep means that the approximate final temperature is now being displayed; the temperature will only 

change marginally from this point on. If the temperature is being measured under the arm, it is important to 

stick to the intended minimum measurement, irrespective of whether the device beeps.

• Do not drop the thermometer. It is not resistant to shock or impact.

• The thermometer is not completely waterproof and therefore may only be immersed into water or desin-

fectant up to the nameplate on the back.

• This thermometer is only intended to be used to measure human body temperature.

• The thermometer contains small parts (battery etc.) which children could swallow. Do not let children use 

the device unsupervised.

• Protect the device from high temperatures and direct sunlight.

• Clean the thermometer before and after each use.

• Do not open the casing (except to change the battery).

• Before use, check that the tip and the display are not damaged.

• Do not walk, run or speak while using the thermometer.

• Store the thermometer with put-on protection cap when not in use.

• Using the thermometer other than as intended can lead to inaccurate measurements.

CLASSIFICATION OF THE TEMPERATURES MEASURED

On average, body temperature measured orally is 36.7 °C ± 0,5 °C. You can use the guidelines below to 

classify the level of fever:

Temperatures measured rectally are generally 0.5 °C higher than those measured orally, while temperatures 

measured under the arm are 0.5 °C lower.

Warning: Always consult a doctor if the reading shows a high fever.

37.5 °C to 37.9 °C: elevated temperature 

38.0 °C to 38.9 °C: moderate fever 

39.0 °C and above: high fever

CHANGING THE BATTERY

When the display shows the   symbol, the battery needs to be changed.

• Remove the screw on the back of the thermometer. Remove the battery compartment cover from the 

thermometer.

• Do not use a metal object to push the battery out of its holder.

• Carefully pull the battery holder about 1 cm out of the casing.

• Insert a new battery (LR41 or SR41, 1.55V) with the + sign towards the top.

• Push the battery holder back into the casing. Replace the battery compartment cover and secure it with 

the screw.

METROLOGICAL INSPECTION

TECHNICAL DETAILS

WARRANTY

DISPOSAL

GENERAL PROVISIONS

CLEANING

Basically, a metrological inspection is recommended at intervals of 2 years. However, professional users 

in Germany have to comply with the aforementioned according to „Regulation for Operators of Medical 

Devices“. This can be implemented either by UEBE Medical GmbH, an authority responsible for metrology, or 

authorised maintenance services. For this, please observe your national provisions.

© 2013 UEBE Medical GmbH, Not to be reproduced, either in whole or in 

part, subject to technical modifications

7 0833 001 C  Mrz 13

Type:

Maximum thermometer, direct mode

Range:

32.0 °C to 42.9 °C 

„L“ for low is displayed for temperatures below 32.0 °C, „H“ for high is dis-

played for temperatures over 42.9 °C

Accuracy:

± 0.1 °C between 35.5 °C and 42.0 °C 

± 0.2 °C in other temperature ranges

Operating conditions: Ambient temperature 10 to 40 °C, relative humidity 30 to 85 %
Storage and transport 

conditions:

Ambient temperature -10 to 60 °C, relative humidity 25 to 90 %

Battery:

LR41 or SR41, 1,55 V

IP rating:

IP 20: protection from solid foreign bodies with diameters of 12.5mm and above; 

no protection against water

Batteries and technical appliances must not be disposed of with domestic waste, but should be hand-

ed in at the appropriate collection and disposal points.

Clean the thermometer before and after every use, using a soft cloth and isopropyl alcohol diluted with 

water, or cold soapy water.

The thermometer is not completely waterproof and therefore may only be immersed into water or desin-

fectant up to the nameplate on the back. Do not use boiling water, gas or a steam autoclave to sterilise 

the device.

The device has been manufactured and tested with great care. However, in the unlikely event of a defect 

being detected after delivery, we provide warranty in accordance with the following terms and conditions:

This product complies with the Council Directive 93/42/EC from 5 September 2007 regarding med-

ical devices, which became effective on 21 March 2010 and bears the mark CE 0123 (TÜV SÜD 

Product Service GmbH).

Storage and transportation conditions  

Ambient temperature -10 to 60 °C, rel. air humidity 25 to 90 %

Degree of protection against electric shock: TYP BF

Please observe operating instructions

Lot number

Keep dry

Manufacturer

During the warranty period of 2 years from the date of purchase we reserve the right either to repair any 

such defect at our expense or to supply a perfect replacement unit. The cost of returning the unit to our 

factory shall be borne by the sender. UEBE shall refuse to accept return deliveries that have not been paid 

for by the sender.

Excluded from the warranty are parts subject to normal wear and tear as well as damage caused by 

non-compliance with the instructions for use, improper handling (e.g. unsuitable power sources, breakages, 

leaking batteries) and/or disassembly of the unit by the purchaser. Furthermore, no claims for damages 

against us are substantiated by the warranty.

Warranty claims can only be advanced in the warranty period and by presenting proof of purchase. In the 

event of a warranty claim, the unit must be sent to the following address together with the proof of purchase 

and a description of the complaint: UEBE Medical GmbH, Zum Ottersberg 9, 97877 Wertheim, Germany

In the case of defectiveness of the goods, the contractual rights of the purchaser to claim against the seller 

in accordance with § 437 German Civil Code are not limited by the warranty.

Please note: In the event of a warranty claim it is essential to attach the proof of purchase.

UEBE Medical GmbH 

Zum Ottersberg 9 

97877 Wertheim 

Germany

Phone +49 (0) 9342 924040  

Fax +49 (0) 9342 924080  

E-Mail [email protected]  

www.uebe.com

60 °C

-10 °C

90%

25%

Отзывы: