15
5. Settings and calibration
5.1 Drying Mode setting
1. Standard drying mode
This mode is the default setting and is suitable for most samples. The
heating temperature can be set, the sample is heated to the set
temperature during the test, and the temperature is maintained until
the test ends automatically.
2. Fast drying mode
This mode is suitable for samples with high moisture content. The
heating temperature can be set, and the initial temperature will exceed
the set value, and will later drop to the set temperature and remain until
the test ends automatically.
3. Soft drying mode
This mode is suitable for samples with low moisture content. The time
required for the heater to rise to the set temperature is longer, and the
temperature rises evenly and slowly. The time during the temperature
rise phase can be set manually.