UNIT MENU
This menu activates units so they will be accessible with
the
UNITS
button. The units in the menu must be turned
on to be active. (Default settings are indicated in bold.)
NOTE: Available units vary by model and local
regulations. The security switch must be in the
unlocked position to turn units on and off.
g
off,
on
lb:oz
off,
on
kg
off,
on
grn
off,
on
ct
off,
on
hkt
off,
on
N
off,
on
sgt
off,
on
oz
off,
on
twt
off,
on
ozt
off,
on
t
off,
tola, tical
dwt
off,
on
C
off,
on
lb
off,
on
End Unit
exit menu
CUSTOM UNIT (C)
Use the custom unit to display weight in an alternative
unit of measure. The custom unit is defined using a
conversion factor based on gram unit, where the
conversion factor is the number of custom units per
gram expressed in scientific notation (Factor x 10
n
).
FACTOR
Set the conversion factor (0.1 to 1.99) using the numeric
keypad. The default setting is 1.0.
EXPONENT
Set the factor multiplier.
-3 = divide the Factor by 1000 (1 x 10
-3
)
-2 = divide the Factor by 100 (1 x 10
-2
)
-1 = divide the Factor by 10 (1 x 10
-1
)
0 = multiply the Factor by 1 (1 x 10
0
)
1 = multiply the Factor by 10 (1 x 10
1
)
2 = multiply the Factor by 100 (1 x 10
2
)
3 = multiply the Factor by 1000 (1 x 10
3
)
LEAST SIGNIFICANT DIGIT
Set the graduation.
Settings of 0.5, 1, 2, 5, 10, 100 are available.
NOTE: Custom unit is locked at the off position
when security switch is set to the locked
position.
Follow instructions below to enter the unit menu to set
the custom unit.
Press and hold
MENU
until
MeNU
is displayed. When
the button is released, the display will show
C.A.L
.. Press
NO
through the succeeding menu items:
C.A.L
.,
SEtUP
,
ModE
until
Unit
is displayed.
1. When
Unit
is displayed, press
YES
.
2.
rESEt
will be displayed. Press No,
Unit g
will be
displayed.
3. Press
NO
to toggle through active units until
Unit c
is
displayed.
4. Press
YES
,
On
c
will be displayed, then press
YES
.
5.
FActor
is displayed. Press
YES
to view the default
Factor. Press
YES
to accept or
NO
to edit the "factor
multiplier". The stored value is then displayed with
the first digit highlighted
1
.0000000
. Repeatedly
press
NO
until the desired number appears. Press
YES
to accept and highlight the next digit. Repeat
until all the digits are correct. Press
YES
to accept the
"Factor" value.
6.
EPt
is displayed. Repeat the same procedure as in
step 5 to accept or edit the "Exponent" value. Press
YES
to accept the "Exponent" value.
7.
LSd
is displayed. Repeat the same procedure as in
step 5 to accept or edit the "Least Significant Digit"
value. Press the
YES
key to accept the graduation
value,
End
is displayed.
8. When
End
is displayed, press
EXIT
to exit to
weighing mode.