0113 TH-2382
PAGE 2 OF 2
SCOUT PRO BALANCE SCALE
H-2382, H-2383
H-2384
ERROR CODES
ERROR CODE LiST
RECOMMENDATiONS
Err 1 chEct
Invalid checksum data.
1. Cycle balance ON/OFF
2. Return balance for service
Err 2 LoAd
Overload (>cap+90d) or Underload.
1. Remove load
2. Make sure shipping lock is unlocked
3. Recalibrate
Err 4 WEight
Incorrect calibration weight.
1. Use correct calibration weights
2. Check environmental conditions
Err 5 999999
Displayed value>99999.
1. Exit application
2. Rezero balance
Err 6 count
Parts counting error – balance exits parts
counting.
1. Exit application
2. Rezero balance
Err 8 FULL
RS232 buffer is full.
1. Cycle ON/OFF
2. Check RS-232 settings in balance and computer
Err 9 dAtA
Internal data error. The next button
press or a 5 second time-out causes the
balance to turn off.
1. Cycle ON/OFF
2. Return balance for servicing