background image

0113 TH-2382

PAGE 2 OF 2

SCOUT PRO BALANCE SCALE

H-2382, H-2383 
H-2384

ERROR CODES

ERROR CODE LiST

RECOMMENDATiONS

Err 1 chEct 

Invalid checksum data.

1.  Cycle balance ON/OFF

2.  Return balance for service

Err 2 LoAd

Overload (>cap+90d) or Underload.

1.  Remove load

2.  Make sure shipping lock is unlocked

3.  Recalibrate

Err 4 WEight 

Incorrect calibration weight.

1.  Use correct calibration weights

2.  Check environmental conditions

Err 5 999999 

Displayed value>99999.

1.  Exit application

2.  Rezero balance

Err 6 count 

Parts counting error – balance exits parts 

counting.

1.  Exit application

2.  Rezero balance

Err 8 FULL 

RS232 buffer is full.

1.  Cycle ON/OFF

2.  Check RS-232 settings in balance and computer

Err 9 dAtA 

Internal data error.  The next button  

press or a 5 second time-out causes the     

balance to turn off.

1.  Cycle ON/OFF

2.  Return balance for servicing

1-800-295-5510

uline.com

π

  

CHICAGO • ATLANTA • DALLAS • LOS ANGELES • MINNEAPOLIS • NYC/PHILA • SEATTLE • MEXICO • CANADA