Control Solution
(Fig.47)
The Control Solution is to ensure that Tyson Bio HS200-B Blood Glucose Monitoring System is working properly and
you are performing a test correctly.
When to Perform a Control Solution Test
1. Be sure of the Control Solution is at a room temperature between 59°F ~ 95°F (15°C ~ 35°C)
before testing.
Control solution that is not at room temperature when tested may cause out of range results.
2. Insert a Tyson Bio 50H Blood Glucose Test Strip, top side up, contact bar end first, into the test
port. The meter will be automatically turned ON (Fig.47).
3. A beep will sound and the system checking will appear followed by a flashing “ ” icon (Fig.48).
Performing a Control Solution Test
1. You use the meter for the first time.
2. You use a new vial of test strips.
3. You suspect that the meter or test
strips are not working properly.
4. Your test results are not consistent with how you feel.
5. You think your test result is not accurate.
6. You damage your meter.
7. An advice from your healthcare professional.
Note: Please read the Control Solution package insert for more detail information.
(Fig.48)
23