Page 92
12.3
Draining the Reservoir
Caution!
Before starting, Make sure that the electric cord is disconnected
and there is no pressure in the autoclave.
The drain valve is located on the front left side of the autoclave after
the door is opened. The function of the drain valve is to drain the water
reservoir.
1.
Connect the silicone hose, supplied with the autoclave, to drain into a
bucket.
2.
Turn drain valve counterclockwise to the open position.
3.
Fully drain the reservoir.
4.
With a quart of tap water flush out the reservoir.
5.
Turn drain valve clockwise to the close position.
6.
Connect the electric cord to power source.
7.
Fill the reservoir with distilled water to just below the safety valve (see
sec. 9.1).
8.
Turn on the main power switch.
9.
The autoclave is now ready for use.