Page 20
3.9
Utilities
3.9.1
Electrical utility
Property
Value: LABSCI 15L
Value: LABSCI 15+L
Power supply
208/230VAC ±10%, 50Hz 400VAC ±10%, 50Hz
Recommended circuit
breaker
15A
10A
3.9.2
Other Utilities
Compressed Air (C
models only)
1/2" 3 Bar (44-58 psi)
Tap water
1/2", 2-6 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp. of 80°C
(176°F)
Attention:
•
A switch or circuit-breaker must be included in the building installation.
This switch or circuit-breaker shall be in close proximity to the
equipment, within easy reach of the operator; and marked as the
disconnecting device for the equipment.
•
The electrical net must be protected with a current leakage safety
relay.
•
The electrical network must comply with local rules or regulations.
•
Verify that there is an easy access to the main power switch, to the
water cut-off valve and to the current leakage safety relay.
•
Make sure while placing the autoclave, to leave space around the
machine, to give the technician access to service the machine.
•
All water connections to autoclave must be performed through "BACK
FLOW PREVENTION SYSTEM" only, as per IEC 61770.