Page 44
Cautions!
Both PT100 temperature sensors must be inside the bottles.
For proper sterilization, fill the bottles with approximately the
same amount of liquid.
Nominal parameters default settings
•
Sterilization temperature: 121
C (250
F)
•
Sterilization time: 45 minutes
Operations sequence:
•
(PV, pre-vacuum models only)
Pulse low/Pulse high:
At one
pulse it will build vacuum down to 25 kPa.
All exhaust from the
chamber is performed through the bio-hazard filter.
•
Water Inlet
: Water enters the chamber
•
Heating:
The chamber is heated by actuation of electrical heaters
until the sterilization temperature is reached.
•
Sterilization:
Sterilization temperature is maintained constant during
the sterilization time.
•
Cooling:
N/A
•
Slow Exhaust:
Steam is exhausted from the chamber at a slow rate,
until it reaches the pressure of 30 kPa above the ambient pressure.
Then steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
Note:
If the cycle fails, fast exhaust is performed through the bio-hazard
filter. Some water may remain in the chamber after a failed cycle. To
remove such water, press Start and repeat the cycle. If the problem
persists, call for service.
•
Drying:
N/A
•
Ending:
The system checks that End temperature and End pressure
have been reached, and then releases the door.
5.10
Program 10: Vacuum Test (PV, pre-vacuum models
Содержание LabSci 11-V
Страница 19: ...Page 18 8 Right service door lock...
Страница 22: ...Page 21 10 Drain outlet...
Страница 34: ...Page 33 Attention This device is not a medical device and not intended for medical use...
Страница 35: ...Page 34 During the process the stages of the cycle will be displayed on the screen...
Страница 74: ...Page 73 1 Open the printer s cover door 1 by pulling it up see Fig 2 Fig 1...
Страница 79: ...Page 78...
Страница 90: ...Page 89 Attention Use protective gloves in order not to burn your hands with the hot steam Safety valve ring...