When all changes are completed, use the
START/STOP key
to move to SET, then use the
UP or DOWN key
to set the new date and time
(this needs to be done even if the
0 date and time are correct).
•
When saving is completed, the main screen will appear
10. WARNING
:
The
EZ11Plus
is equipped with a
pre-heat mode
for the “Wrapped
Instruments, Pouches”, “Handpieces” and “Custom A” cycles. Upon initial set up after turning
on the power switch and having set the date and time, it is critical to open the door, select
the “Unwrapped Instruments” cycle and empty the chamber of all packing materials.
To do so:
•
Open the door immediately after
System Ready
has flashed across the screen followed
by the fill reservoir icon . Now use the arrow keys on the EZPad to
advance to the “Unwrapped Instruments” cycle.
•
This will turn off the pre heating mode (the EZ9Plus does not have this pre heating
feature).
11. Remove all accessories and plastic material from the chamber.
12. Fill the reservoir, using one of the two options below, with steam distilled water having the
characteristics described in sec 3.8 of the Operator’s Manual.
NOTE:
The reservoir has a built in overflow. Over filling the reservoir will result in water
emptying on to the counter top under unit
Fill from the front funnel as follows:
•
Open the door. (The autoclave needs to be on for the door to open).
•
Pour steam distilled water, gently, into the front funnel until it reaches the top of the blue
area on the level gauge.
•
If water is filled above the blue area into the red area then use the drain hose to drain off
the over fill (see sec 12.2 in the Operator’s Manual).
Fill from the opening at the top as follows:
•
Remove the water reservoir cover.
•
Pour steam distilled water into the reservoir through the opening on top of the autoclave
until it reaches the base of the safety valve holder or reaches the top of the blue area of
the front level gauge.
•
Under no circumstance should water be filled above the safety valve holder (located in the
reservoir).
•
In case more water is accidentally filled above the blue area, decrease the water level by
draining the reservoir before starting a cycle (see sec. 12.2 in Operator’s Manual).
13. Using the tray rack reinsert the trays back into the chamber. In the EZ11Plus it is normal and
necessary that the bottom tray protrude more than the other trays. DO NOT push this tray all
the way in, it will interfere with the sensors at the back of the chamber.
14. The unit is now ready to run sterilization cycles.
This Quick Installation Guide is an excerpt from the Operator’s Manual. The Operator’s Manual should be read in full to gain an
understanding of proper use and maintenance of this autoclave. The Operator’s Manual should also be saved for reference.
EZPLUS Autoclave Quick Installation Guide
Water reservoir cover
Safety valve holder
12a.
12b.
Ster. Temp.
270.0
°F
Ster. Time
4.0
min
Dry Time
30
min
System Ready
Wrapped
Instruments, Pouches
Temperature
109.6
°F
Pressure
00.09
inHg