When all changes are completed, use the

 START/STOP key

 to move to SET, then use the 

 

UP or DOWN key

 to set the new date and time 

(this needs to be done even if the 

 0 date and time are correct). 

• 

When saving is completed, the main screen will appear

10.  WARNING

:

 The 

EZ11Plus

 is equipped with a 

pre-heat mode

 for the “Wrapped 

Instruments, Pouches”, “Handpieces” and “Custom A” cycles. Upon initial set up after turning 
on the power switch and having set the date and time, it is critical to open the door, select 
the “Unwrapped Instruments” cycle and empty the chamber of all packing materials.  
To do so: 

•  

Open the door immediately after 

System Ready

 has flashed across the screen followed  

  by the fill reservoir icon        . Now use the arrow keys on the EZPad to 
  advance to the “Unwrapped Instruments” cycle. 

 

• 

This will turn off the pre heating mode (the EZ9Plus does not have this pre heating 

  feature). 

11. Remove all accessories and plastic material from the chamber.

12. Fill the reservoir, using one of the two options below, with steam distilled water having the 

characteristics described in sec 3.8 of the Operator’s Manual. 

NOTE:

 The reservoir has a built in overflow. Over filling the reservoir will result in water 

emptying on to the counter top under unit 
 
Fill from the front funnel as follows: 

 Open the door. (The autoclave needs to be on for the door to open).  

 Pour steam distilled water, gently, into the front funnel until it reaches the top of the blue  

  area on the level gauge.  

 If water is filled above the blue area into the red area then use the drain hose to drain off  

  the over fill (see sec 12.2 in the Operator’s Manual).  

 
Fill from the opening at the top as follows: 

 Remove the water reservoir cover. 

 Pour steam distilled water into the reservoir through the opening on top of the autoclave  

  until it reaches the base of the safety valve holder or reaches the top of the blue area of  
  the front level gauge. 

 Under no circumstance should water be filled above the safety valve holder (located in the  

  reservoir). 

•  

In case more water is accidentally filled above the blue area, decrease the water level by 

  draining the reservoir before starting a cycle (see sec. 12.2 in Operator’s Manual).

13. Using the tray rack reinsert the trays back into the chamber. In the EZ11Plus it is normal and 

necessary that the bottom tray protrude more than the other trays. DO NOT push this tray all 
the way in, it will interfere with the sensors at the back of the chamber.

14.  The unit is now ready to run sterilization cycles.

This Quick Installation Guide is an excerpt from the Operator’s Manual. The Operator’s Manual should be read in full to gain an 
understanding of proper use and maintenance of this autoclave. The Operator’s Manual should also be saved for reference.

EZPLUS Autoclave Quick Installation Guide

Water reservoir cover

Safety valve holder

12a.

12b.

Ster. Temp.

 

270.0 

°F

Ster. Time

 

4.0   

min

Dry Time

 

30    

min

System Ready

Wrapped

Instruments, Pouches

Temperature

109.6

 

°F

Pressure

00.09

 

inHg