EZPLUS Autoclave Quick Installation Guide

1.  Examine the outer carton and autoclave for any signs of damage. Immediately notify your 

dealer or Tuttnauer USA of any signs of damage.

2.  To avoid injuries, lifting and carrying should be done by two people. 

  Lifting straps have been installed for your convenience. 

Lifting straps are for 

one time 

use 

only and should be removed and discarded after initial set up.

3.  Place the sterilizer on a rigid level surface. The counter top or stand must be able to 

support  

 

175 for the EZ9Plus 

 

190 for the EZ11Plus

4.  The minimum depth of the counter top needs to be 22 inches. 

5.  Leave a minimum 2” (50mm) space around the back and sides of the unit. 

6.  Make sure all the feet are on the autoclave and none of them have been lost.

7.  Connect the power cord to the socket on the rear of the autoclave; then plug it into the 

supply outlet. 
  a.  The supply outlet must be a properly grounded outlet.  
  b.  It is mandatory to have  installed an earth leakage relay (GFCI outlet or GFCI  
      circuit breaker). 
  c. The voltage supplied to the  autoclave must comply with the voltage requirements 
      stipulated on the device label ±5% (the device label is affixed to the side of the 
       autoclave). 
  d. It is recommended that the  device be on a dedicatedelectrical circuit.

8.  The EZPlus is equipped with an electronic door lock. The door will be remain locked if: 

  a.  The power is off 
  b.  The EZPlus is running a cycle and the cycle has not completed.  
  c.  An error occurs (this is cleared by pressing the STOP key) 
  d. There is water in the chamber. 

9.  Turn on the power switch, located on the right side of the unit. A message “Set Date and 

Time” will appear on the screen. (On initial set up, the date and time MUST be set to turn 
on the internal clock in the unit. Failure to do this will result in a “Time Error”) To set the 
date and time follow these steps:

 

• 

Use the 

UP

 and 

DOWN

 keys to change the blinking digit.

 

  Use the 

START/STOP

 key to move to the next digit. 

• 

The time is displayed in the upper row in the form “HH:MM”.  

• 

The time is in a 24 hour format (i.e. 2:30PM = 14:30) 

  The date is displayed in the lower row in the form “DD/MMM/YYYY” (e.g. 05/ 

  APR/2012) 

Power switch

Minimum

22”depth

counter top

required

lbs

S i n ce

1925

Tuttnauer USA., Co., Ltd.

25 Power Drive, Hauppauge, NY 11788. Tel: +800 624 5836, +631 737 4850 Fax: +631 737 0720
E-mail: [email protected], www.tuttnauerUSA.com