Page 10
Read the Operating Instructions carefully, before beginning any
operation on the autoclave!
1.10.
Operating Conditions
This device is for indoor use only!
The minimum counter depth for the Elara11 is 24 inches.
Counter tops or slide-outs need to have a minimum capacity of 275 pounds.
The sterilizer should be loaded only with autoclavable material!
Minimum room ventilation shall be 10 cycles per hour.
The environment shall not exceed an ambient temperature range of from 41ºF
(5ºC) to 104ºF (40ºC) and a relative humidity of 85%.
Operate the autoclave only in the manner specified in the manual. If the
equipment is used in a manner not specified by the manufacturer, the
protection provided by the equipment may be impaired.
CAUTION!
Wastewater should be brought into the public net in
accordance with the local rules or requirements
ONLY NON-HAZARDOUS LIQUIDS SHALL BE DISPOSED IN
PUBLIC SEWAGE!