9.3
Monthly Maintenance
Caution!
Make sure the autoclave is not hot before cleaning it.
l
Turn the unit on momentarily to allow the door to be opened. Open the
door, unplug the autoclave again, and proceed with cleaning.
l
Clean and descale the chamber.
l
If the autoclave is only used occasionally, drain the water from the mineral
free water reservoir once a week, and refill with fresh mineral-free water or
distilled water.
l
Once a week or if a text alarm of ‘full waste water reservoir’ appears
(whichever comes first) drain the water from the waste water reservoir.
l
Clean the outer parts of the autoclave with a soft cloth.
l
Clean the Drain filter of the autoclave.
9.4
Periodic Maintenance
Caution!
Make sure the autoclave is not hot before cleaning it
1. Replace the air filter, every 6 months or after 1000 cycles
(whichever comes first) according to section 12.2.
2. Every 3 months check the door gasket for any signs of physical
damage and ask the technician to replace it if there is tear or
leakage.
9.4.1
Replacing the Air Filter
Cautions!
Before proceeding, make sure that the electric cord
is disconnected and there is no pressure in the
chamber.
Use scissors to open the filter bag and not sharp
blades or pointed instrument.
Check that the new filter has not exceeded the
maximum shelf life.
Carefully un-pack the new filter and examine it for
any signs of damage.
Remove any protective packaging before inserting
Page 64
Содержание AMS10-120-T
Страница 2: ......
Страница 4: ...https www tuttnauerusa com...
Страница 16: ...1 10 Device Specifications 1 10 1 Device Overall Dimensions Page 9...
Страница 55: ...Note For the results of the cycle see description in section 8 5 Page 48...
Страница 66: ...Press to receive the entire printout of the selected cycle history Page 59...
Страница 85: ...Page 78...
Страница 104: ...Cat No MAN205 0502015EN Rev H COPYRIGHT...