ASSEMBLY & DISASSEMBLY

3

MAINTENANCE

REPAIR

1. Determine direction for fluid to flow.

2. Install meter observing directional arrow on casting.

3. Remove four screws (item 28).

4. Rotate meter cover assembly (item 37) to desired

orientation.

5. Replace four screws.

Meter consists of a chamber housing, measuring chamber,

gear train, counter assembly and cover.  Meter can be com-

pletely disassembled without disturbing piping, or meter can

be partially disassembled as required.

Counter Assembly

For access to counter assembly, remove reset knob (item 15)

by grasping edges and pulling firmly.  Knob is held in place by

a spring clip.  Loosen two screws (item 14) and lift bezel off.

Remove two screws (item 12) to detach counter face (item 13).

Remove two screws (item 9) to extract counter (item 10).

Reassemble by reversing procedure.

Meter Chamber Assembly

To expose meter chamber assembly, gear train and seal,

remove four screws (item 28).  Meter chamber assembly

consists of upper and lower chambers, a nutating disc and four

screws.  Meter chamber assembly (item 3) can be dislodged

by removing four screws (item 5).  Reassemble by reversing

procedure.

If replacement of any components of the meter chamber

assembly is required, the complete assembly must be

replaced due to the precise method of its construction.  This

assures a proper fit and a correctly operating chamber.

Gear Train and Seal

To disassemble gear train and seal, remove two screws (item

5) and gear frame (item 6).  Remove cluster gear (item 18) and

washer (item 19).  Remove drive gear (item 24) and washers

(item 23) by rotating and pulling drive gear.  Remove O-ring

seal (item 25).

When reassembling seal, lubricate O-ring with oil or petroleum

jelly and replace in cover.  Place washer on drive gear shaft.

Rotate and push shaft through O-ring and cover carefully to

prevent damage to O-ring.  Shaft must then be guided into

pinion bevel (item 27) if counter has not been removed.

Replace remaining parts to complete assembly by reversing

disassembly procedure.

The Sotera Systems Series 900 meters can be calibrated for

either U.S. gallons or liters.  Calibration is required upon

installation, after disassembly, after significant wear or when

metering a different viscosity fluid.  Depending on the model,

Series 900 meters are calibrated at the factory metering

Stoddard solvent in either U.S. gallons or liters.

Meter calibration can be easily changed by following the

calibration procedure listed below.  A proving container or a

container of KNOWN volume will be needed for the calibration

procedure.  It is recommended that the container's volume be

at least five times larger than the unit of calibration.  For example,

a five gallon container should be used when calibrating for

gallons.

Procedure for Calibration:

1. Fill container to a known volume.

2. If indicated amount does not match known volume, turn

calibration screw (item 31) counterclockwise to reduce

indicated amount or clockwise to increase the amount.

3. Repeat step 2 until calibration is acceptable.

For accurate measurement and to prevent meter damage,

meter and piping must always be filled with liquid and free of air.

Meter should be calibrated per instructions in this manual prior

to its use.

1. Reset meter to “0”.

2. Meter is ready for use.

Meter should operate maintenance free.  However, certain

liquids can dry out while in meter housing, causing meter to stop.

If this happens, meter should be thoroughly cleaned (see

instructions below).

Cleaning Instructions:

Run a flushing fluid through meter.  For a more thorough

cleaning, disassemble meter per “ASSEMBLY/

DISASSEMBLY” section, “Meter Chamber Assembly”

subsection.  Rinse all meter components.   Recalibrate meter

following calibration instructions above.

Storage:

If meter is to be stored for a period of time, clean thoroughly.  This

will help protect meter from damage.

Meters needing repair should be taken to an authorized repair

shop or returned to factory for service.  Meters must be

thoroughly triple-rinsed before being taken in for repair.

CALIBRATION

OPERATING INSTRUCTIONS

Содержание Sotera 900 Series

Страница 1: ...1 1 2 911TL 1 1 2 Owner s Operation Safety Manual OUTSTANDING FEATURES 6 to 40 GPM 23 to 151 LPM flow rate Accurate to within 2 1 or 1 1 2 NPT inlet outlet ports Large easy to read numbers Quick reset knob Accumulative totalizer Teflon coated aluminum body ...

Страница 2: ...ounter has four unit wheels which can be reset to zero It s totalizer has eight unit wheels INSTALLATION QUALITY OPTIONS 2 GENERAL DESCRIPTION TECHNICAL INFORMATION Dear Sotera Customer Thank you for buying a Sotera product Sotera Systems represents a new age in chemical handling products This service guide contains valuable information about your new equipment and its operating and service requir...

Страница 3: ...e remaining parts to complete assembly by reversing disassemblyprocedure The Sotera Systems Series 900 meters can be calibrated for either U S gallons or liters Calibration is required upon installation after disassembly after significant wear or when metering a different viscosity fluid Depending on the model Series 900 meters are calibrated at the factory metering Stoddard solvent in either U S ...

Страница 4: ...ot adhering to these specifications may berefusedserviceateitherthe repair shop or at the factory ITM PART PROC NO NO NO DESCRIPTION QTY 7 0HWHU RXVLQJ 7HIORQ RDWHG 9 DVNHW OXRURFDUERQ 0HWHU KDP EHU VVHP EO 9 RYHU DVNHW OXRURFDUERQ 3 06 5 7 HDU UDP H 7 0HWHU RYHU 7HIORQ RDWHG 3 06 7 SH 3 06 5 77 RXQWHU VVHP EO 8 6 DOORQ RXQWHU VVHP EO LWHU 2SW H HO 3 06 RXQWHU DFH RXQWHU DFH LWHU 2SW 2 06 7 SH QRE...

Страница 5: ...uring chamber or gears Clean or replace internal metering sticking components Shaft seal leakage Dirty seal Clean O ring seal and seat area Bad seal Replace seal Gasket leakage Loose joints Tighten joints Dirty gasket Clean gasket and seat area Bad gasket Replace gasket Low flow capacity Clogged meter chamber Clean meter chamber TROUBLESHOOTING GUIDE ...

Страница 6: ...ems Tuthill Corporation has ushered in a new age in chemical handling products for the ag chem dealer and agricultural user After more than 35 years of product research and development in fuel and chemical handling equipment Tuthill Corporation has created Sotera Systems to focus solely on the needs of the ag chemindustry You ll recognize the familiar Tuthill traits in every Sotera Systems product...

Отзывы: