#3 Calibration with Your Standard
Now that the blank is set, you should calibrate with
your standard.
1. Access screen 2.0
2. Press <2> to access screen 2.2 and input the actual
concentration of your standard. Return to screen 2.0
3. Insert and cover your standard.