LCD ERROR MESSAGES 

The thermometer incorporates visual diagnostic messages as follows: 

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of HAL and 
LAL. 

 

 

‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient temperature. ‘Er3’ 
is displayed when the ambient temperature exceeds 0

o

C (32°F) or +50

o

C (122°F). The thermometer 

should be allowed plenty of time (minimum 30 minutes) to stabilize to the working/room temperature. 

 

Error 5~9

, for all other error messages it is necessary to reset the thermometer. To reset it, turn the 

instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery and turn 
on. If the error message remains please contact the Service Department for further assistance. 

」「

 

‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the measurement range.

 

BATTERIES

 

The thermometer incorporates visual low battery indication as follows: 

 

 

 

‘Battery OK’: measurements 

 are possible 

‘Battery Low’: battery needs to be  

replaced, measurements are still possible 

‘Battery Exhausted’: measurements are not 

possible 

 When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries. Please note: 

It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction. 

 Dispose of used battery promptly and keep away from children. 

SPECIFICATION 

Item 

Non-contact Infrared Scan function 

Thermocouple 

Probe 

Scan 

function 

(K type; probe not included.) 

Measurement Range 

-60 to +625

 o

C (-76 to +1157

 o

F)

 

-64 to +1400°C (-83.2 to +1999°F) 

Operating Range 

0 to +50°C (32 to +122°F) 

Accuracy  
(Tobj=15~35°C, Tamb=25°C) 

±1.0°C (1.8°F) 

+/-1%  of  reading  or  1°C  (1.8°F)  whichever  is 
greater (Test under Tamb=23±6

°

C) 

Accuracy(Tamb=23±3°C) 

Tobj=0~625°C:+/-2%  of  reading  or  2°C  (4°F)  whichever  is 

greater, Tobj= -60~0

 o

C: +/-(2°C+0.05/degree)

 

Emissivity Range 

0.95 default – adjustable 0.1 to 1 step .01 

Resolution (-9.9~199.9°C) 

0.1°C/0.1°F 

Response Time (90%) 

1sec 

Distance:Spot 

16:1(90% energy covered) 

Battery Life 

Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.) 

Dimensions 

46.0 x 143.0 x 184.8mm(1.81×5.63×7.28 inch) 

Weight 

240 grams(8.5 oz) including batteries (AAA*2pcs) 

Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10

 o

C (18

 o

F). 

 

 