1) Remove the 4 screws

2) Open the bottom lid being careful 

not to stretch the connected wires.

3) Remove the O-ring

6) Slide in the new CR2477 battery 

into the holder, with the "+" face up

4) Lift the PCB by holding 

the battery holder

5) Slide the battery out of the battery holder by pushing 

it from the side using your finger.  

Do not

 use a conductive 

object such as a metal screw driver.

PROCEDURE FOR

 

BATTERY CHANGE

3/5

The latest Vitalbase provides the option for changing the battery. 

To change the battery please carefully follow the below steps

Содержание Vitalbase

Страница 1: ...for 3 seconds The red led lights up Immediately release the button 1 2 3 Push and release the button The red led turns on then turns off VitalBase is ready LED Alarm Button After removing VitalBase from its package fully read the documentation before performing the following steps For optimal results VitalBase has to be permanently worn on the wrist Other uses may generate false alarms Release5 Ja...

Страница 2: ...e watch back into sleep mode for transit or storage please follow the below process 2 5 The VitalBase watch is designed to identify a serious fall that leads to a state of immobility and or unconsciousness of the user If such a fall is detected the detector vibrates to alert the user that it is about to send a radio alarm signal to the Lifeline home unit Prior to the unit vibrating in order to red...

Страница 3: ...the face up 4 Lift the PCB by holding the battery holder 5 Slide the battery out of the battery holder by pushing it from the side using your finger Do not use a conductive object such as a metal screw driver PROCEDURE FOR BATTERY CHANGE 3 5 The latest Vitalbase provides the option for changing the battery To change the battery please carefully follow the below steps ...

Страница 4: ... aligning the small vertical mark on the inside of the case with the dark straight line on the PCB 9 Place a new O ring into the case top circular groove 4 5 Contact Tunstall on 01977 660479 to obtain a new O ring and Vitalbase case pack ...

Страница 5: ...the wires behind the lid s internal plastic leg located behind the vibration motor 11 Make sure the vibration motor wires are not pinched when closing the bottom lid 5 5 13 Test the Vitabase by pressing the help button and ensure an alarm is generated to the monitoring centre Tell the monitoring centre you are testing the Vitalbase following a battery change ...

Отзывы: