Tunstall iVi Скачать руководство пользователя страница 12

 

Menu 5 – Not Worn Period 

Flashing Red LED 

When the Not Worn alert is active (See Menu 2 – Advanced Mode), the number of days 
before the iVi will send a Not Worn event can be changed using this menu.  
 
Pressing the Help button will cycle between enabling/disabling the button. After each press a 
different tone will sound to indicate the setting selected. When the correct setting is selected, 
either press the cancel button to move to Menu 6 or press no further buttons and after 20 
seconds the iVi will exit the programming menu.  

Option 

Setting 

Indicator Tone 

Note 

7 Days 

High Tone 

Default setting 

5 Days 

Medium Tone 

 

3 Days 

Low Tone 

 

 

 

Menu 6 – Sleep Mode (for transportation) 

Flashing Orange LED 

To stop the iVi from generating radio transmissions (and beeps etc) during delivery and for 
storage, the unit is supplied in a sleep mode. When in sleep mode, a press of the Help 
button will exit sleep mode and return it to normal operation. A subsequent press of the Help 
button will generate an alarm call. Sleep mode can be entered again using this menu to 
allow shipment of the device after configuration. 
 
Pressing and releasing the Help button puts the iVi into sleep mode and turns off the LEDs. 
If you don’t want to enter sleep mode, press the cancel button to move to Menu 1 or press 
no further buttons and after 20 seconds the iVi will exit the programming menu.  
 

Help button press = sleep mode   

Cancel button press = move to next menu 

(menu 1) 

 

 

 

 

 

 

 

 

 

 

Press 
Help 
button

 

Содержание iVi

Страница 1: ...iVi the intelligent pendant with auto fall detection Programming and installation Guide...

Страница 2: ...1 Menu 4 Disable Cancel Button 11 Menu 5 Not Worn Period 12 Menu 6 Sleep Mode for transportation 12 Programming the iVi to a Lifeline home unit 13 Using the iVi to answer an incoming call 13 Using wit...

Страница 3: ...this takes up to 20 seconds then it will alert the user by emitting tones from the sounder and flashing the green LED The user then has 10 seconds to cancel the alert by pressing the cancel button The...

Страница 4: ...or Button User has pressed the Help button Call presented to an operator at PNC Code 101 Fall Detector Fall A fall has been detected Call presented to an operator at PNC Code 572 Fall Detector Activat...

Страница 5: ...CareAssist Lifeline 400 4000 Connect Connect Vi and Vi Is the user capable of using the cancel button correctly Once the above points are known it is possible to program the iVi correctly and then pr...

Страница 6: ...f activity Equipment type Normal mode Not worn and Activation cancelled events are turned off Wearing option Neck and Brooch users 5ft 2in or above Waist Worn users 5ft 10in or above If the height of...

Страница 7: ...ogramming mode is accessed by pressing the buttons in the following sequence Press and hold the Cancel button for more than 3 seconds Release the Cancel button the green status LED will light Press an...

Страница 8: ...indicated by the state of the LEDs For more details on each menu please refer to the following pages Menu Number Operation LED Indication Menu options Press alarm button to scroll to next menu option...

Страница 9: ...setting 2 Neck and Brooch 4ft 6in to 5ft 2in Waist Worn users 5ft to 5ft 10in Low Tone Option Setting Indicator Tone Note 1 Cancel Button Enabled High Tone Default setting 2 Cancel Button Disabled Low...

Страница 10: ...for use with all Tunstall home units and radio systems It prevents the Activation Cancelled and Not Worn alerts from being sent by iVi as these alerts are not supported by all systems Advanced Mode M...

Страница 11: ...ndicator Tone Note 1 Neck and Brooch users 5ft 2in or above Waist worn users 5ft 10in or above High Tone Default setting 2 Neck and Brooch users 4ft 6in to 5ft 2in Waist Worn users 5ft to 5ft 10in Low...

Страница 12: ...ow Tone Menu 6 Sleep Mode for transportation Flashing Orange LED To stop the iVi from generating radio transmissions and beeps etc during delivery and for storage the unit is supplied in a sleep mode...

Страница 13: ...nging Once the iVi is programmed to the Lifeline Vi open PC Connect then click on Radio Triggers Events and for the Fall Detector Manual Press event ensure the answering incoming call box is ticked Th...

Страница 14: ...service providers and family members to encourage the user to wear it In order to achieve this the not worn alert can be used in two ways Either Provide an alert to the monitoring centre for an operat...

Страница 15: ...ng and flashing its LED orange once every 30 40 seconds In addition if the error condition persists the iVi will send 3 Automatic Low Battery ALB alerts to the monitoring centre within a 30 minute per...

Страница 16: ...stall does not accept any responsibility for any errors or omissions contained in this document 2016 Tunstall Group Ltd TUNSTALL LIFELINE COMMUNICALL and iVi are trade marks Tunstall Group Limited Whi...

Отзывы: