SWITCHING THE UNIT ON

Press switch On key , at the start Unit Model number & serial number will be
displayed , then pressure range will be displayed

INSTRUCTIONS

Users/customers are expected not to use the instrument below & above the
pressure range mentioned, this will permanently damage the sensor
Always Connect pressure tubes with leak free joints
Instrument will perform “Autozero” automatically for first time at the start &
a message will be displayed
“Autozero successfully done”          
Users can also press zero key manually for 1 second to perform autozero
manually
User can change & select multiple units by pressing unit selection keys 
Upper arrow key /Increment key is used to reset the Minimum & Maximum
values of pressure
Display Back light key is used to switch Off/ On back light , it is
recommended to use backlight when only needed to save battery life ,
backlight automatically goes “switch off” after sometime as programmed in
the setting menu         

Содержание TDP-01B

Страница 1: ...TDP 01B Instrument Manual PressCAL High Accuracy Digital Differential Guage Absolute Pressure Indicator...

Страница 2: ...TABLE OF CONTENTS 1 Preface 3 2 Technical Specification 4 3 Understand the Instruments 5 3 Calibration 9 4 Warranty Certificate 14...

Страница 3: ...2 Operating manual With Warranty certificate 3 Silicon pressure tubes 5 Carrying Bag STANDARED ACCESSORIES Welcome thanks for purchasing Our PressCAL Model No TDP 01B Our high Accuracy pressure measu...

Страница 4: ...f 5 digits For higher ranges depends upon the pressure range Linearity Maximum 0 05 FS Defined as best fit straight line User selectable 12 multiple Pressure measurement units Kpa pa hPa Bar mBar Kg c...

Страница 5: ...tting minimum maximum values Zero Key Enter Press for 1 second to zero tare the pressure display this key works as Enter command in program mode Light Key switch On Off for display light keep display...

Страница 6: ...Battery compartment Remove Silicon protective cover to replace the battery Batteries AA size alkaline...

Страница 7: ...m Autozero automatically for first time at the start a message will be displayed Autozero successfully done Users can also press zero key manually for 1 second to perform autozero manually User can ch...

Страница 8: ...NG INTO SETTING PROGRAM MODE Enter password 1204 with help of shift left key Up arrow Increment key press Zero enter Key to enter the password If no key is pressed for long time the instrument will co...

Страница 9: ...FS throughout the operating range Use this Mode only when you have calibrated external pressure reference source hi accuracy pressure display to calibrate this instrument Connect the External referenc...

Страница 10: ...t XXXX Follow same steps at every 10 till step calibration is complete till SPAN a message will appear that calibration is complete ACTIVATION OF EDIT VALUES IN CALIBRATION MODE Edit Values correction...

Страница 11: ...ion mode with help of enter key After the LCD Display shows the following prompt Edit Table 1 Pressure at Zero 2 Pressure at 10 3 Pressure at 20 4 Pressure at 30 5 Pressure at 40 6 Pressure at 50 7 Pr...

Страница 12: ...left arrow Unit and up arrow key after editing of count value move to edit table with help of exit key light key After Show edit table in display inter Sr no 2 pressure at 10 with help of inter key Ze...

Страница 13: ...key Zero key Repeat the same procedure for other values Edit the count value according to pressure value With help of left arrow Unit and up arrow key after editing of count value move to edit table w...

Страница 14: ...hat Dry Block Temperature Calibrator having Sr No Date is propery Tested for workmanship We Certify our Calibrator for satisfactory performance for the period of one year from the date of supply again...