3. Water-cooled interface installation
1
、
Place the camera on a smooth workbench
2
、
Connect water-cooled pipe to the water pipe connector on the camera and make
sure it is inserted in place as shown below
(
pipe specification
:
inner
①
5mm
,
outerΦ8mm).
3
、
Insert the water pipe into the nozzle of the cooling water circulator and lock it with
clamps.
4
、
Connet the water pipes of cooling water circulator and camera via switching valve.
3.1 Installation procedure of water-cooled pipe
Currently only Dhyana 400BSI V2 and Dhyana 95 V2 offer water-cooled mode
,
Installation schematic diagram is as follows
water-cooled pipe
water connector
- 9 -
- 10 -
3.2 Disconnect procedures of water-cooled pipe
Note
:
After the water pipe is pulled out, a small amount of water remains in
the camera water valve.Please carefully pour the water out with the valve
orifice down.
Water valve sliding sleeve
1
、
Disconnect power for camera and all other equipments including the circulating
water cooler.
2
、
According to the instructions of cooling water circulator, unplug its water pipe
and discharge the water inside.
3
、
Press the slide sleeve of switching valve to pull out the water pipe of cooling
water circulator and discharge the water inside.
4
、
Press the water connector to pull out the camera water pipe. First, turn the side
with installed water valve to the side (not the top). When pulling out the water pipe,
the water valve orifice is facing down. Protect with a absorbent towel or tissue to
ensure no water leaking into the camera (as shown below)