background image

3

Operating Considerations

The TIA-527 is comprised of two 

fi

 ber coupled detectors and two ampli

fi

 er stages. The 

fi

 rst ampli-

fi

 er is a transimpedance stage which converts the net detector output current to a voltage by passing 

it through a resistor of 1 400 or 14 000 ohms.  Additional ampli

fi

 cation is optionally provided by the 

second stage which also serves to provide 50 ohm drive capability. Either AC or DC coupling between 
the stages may be selected.

 The overall bandwidth of the unit is determined by the 

fi

 rst stage transimpedance. It is in excess of 

125 MHz when the T R  switch is in the 1.4 K position and is 35 MHz in the 14 K position.  The over-
all responsivity of the unit in terms of Volts/Watt is the current responsivity of the detector multiplied 
by the transimpedance and further multiplied by the second stage gain. For example, the sensitivity 
of the unit at a wavelength of 1300 nm would be 0.7 A/W x 1400 V/A x 10 = 10, 000V/W. It is evident 
that the same responsivity may be obtained by using a transimpedance of 14 K and a second stage 
gain of one. However, the 

fi

 rst setting will provide a bandwidth of 125 MHz  while the second will pro-

vide a bandwidth of 35 Mhz. Since the overall peak-to-peak output noise increases with bandwidth, 
it it desireable to use the higher transimpedance setting assuming that the signal of interest does not 
exceed 35 MHz.
 
The selection of AC interstage coupling is useful when the user needs to examine a small signal in 
the presence of a large DC optical component, (e.g. baseband 

fi

 ber optic video). This will preclude 

the last stage from saturating on the DC component. Otherwise, DC coupling should be employed.
 
When using the TIA-527 mounted on an oscilloscope, the scope may have its input set to either 50 or 
1 Meg ohm input impedance. If driving a coaxial cable, the cable should have a 50 ohm characteristic 
impedance and be terminated with a 50 ohm load. Note that the signal amplitude will be reduced by a 
factor of two. 

The following table summarizes the operation of the unit under various operating conditions.

Содержание TIA-527

Страница 1: ...A 527 Balanced Optical Electrical Converter Terahertz Technologies Inc 169 Clear Road Oriskany NY 13424 315 736 3642 FAX 315 736 4078 E mail sales terahertztechnologies com 6 2011 Operating Instructio...

Страница 2: ...duction 1 Unpacking and Inspection 1 Battery Replacement 2 Controls 2 Operating Considerations 3 4 Spectral Response 4 Typical Waveforms 5 Specifications 6 Limited Warranty 7 Repair Information 7 Cont...

Страница 3: ...ge in those cases where the signal of interest is combined with a relatively large DC optical component Each unit is powered by a self contained 9 V lithium battery or a universal wall mount power sup...

Страница 4: ...accomplished by removing the four 2 56 Philips flat head screws that retain the bottom cover of the TIA 527 DO NOT attempt to remove the top cover Take care to replace all screws tightly This will pr...

Страница 5: ...10 10 000V W It is evident that the same responsivity may be obtained by using a transimpedance of 14 K and a second stage gain of one However the first setting will provide a bandwidth of 125 MHz whi...

Страница 6: ...X10 Operating Considerations continued Spectral Response The approximate relative response curves of the detectors employed is as shown below Note that these are representative curves and do not nece...

Страница 7: ...5 Typical Waveforms Optical Input Output TR 14 K Gain 1 Optical Input Output TR 1 4 K Gain 10 Optical Input Output TR 1 4 K Gain 1...

Страница 8: ...Frequency Cutoff Output Offset Voltage 0 1 Volt Noise Level 3 6 pW root Hz Maximum Output Voltage 4 V pk pk no load 2 V pk pk with 50 Ohm Load Power Requirements 9 V Battery or supplied universal wall...

Страница 9: ...LIED GIVEN BY TTI I E THERE IS NO WARRANTY OF FITNESS FOR A PARTICULAR PURPOSE TTI HEREBY DISCLAIMS ANY EXPRESS OR IMPLIED WARRANTY OTHER THAN THE WARRANTY IN THE FIRST SENTENCE TO THE FULLEST EXTENT...

Отзывы: