DC Offset Range:
±10V. DC offset plus signal peak limited to ±10V from 50
Ω
/600
Ω
.
DC Offset Accuracy:
typically ±3% ±10mV, unattenuated.
Resolution:
3 digits for both Amplitude and DC Offset.
Pulse Aberrations:
<5% + 2mV.
Aux Out
Multifunction output user definable or automatically selected to be any of the following:
Waveform Sync:
A square wave at the main waveform frequency. Symmetry is 50% for
sine and triangle waves at MAIN OUT; for square waves and pulses
symmetry is the same as that of the waveform at MAIN OUT.
Trigger:
Outputs a replica of the current trigger signal.
Sweep Sync:
Outputs a trigger signal at the start of sweep to synchronize an
oscilloscope. Can additionally output a sweep marker.
Output Signal Level:
Output impedance 50
Ω
nominal. Logic levels of <0.8V & >3V, except
for Sweep Sync. Sweep Sync is a 3-level waveform: low at start of
sweep, high for the duration of the last frequency step at end of sweep,
with a narrow 1V pulse at the marker point.
INPUTS
Ext Trig /Gate
Frequency Range:
DC - 1MHz for FSK; DC – 100kHz for Gate; DC – 2.5kHz for Tone and
Sweep.
Signal Range:
Threshold nominally TTL level; maximum input ±10V.
Minimum Pulse Width:
100ns for Gate and FSK modes; 0.2ms for Sweep and Tone modes.
Input Impedance:
10k
Ω
VCA In
Frequency Range:
DC - 100kHz.
Signal Range:
2.5V for 100% level change at maximum output.
Input Impedance:
typically 6k
Ω
.
INTERFACES
RS-232:
Variable Baud rate, 19200 Baud maximum. 9-pin D-connector.
As well as operating in a conventional RS-232 mode the interface can
be operated in addressable mode whereby up to 32 instruments can be
addressed from one RS-232 port.
USB:
Standard USB hardware connection.
GENERAL
Display:
20 character x 4 row alphanumeric LCD.
Data Entry:
Keyboard selection of mode, waveform etc.; value entry direct by
numeric keys or by rotary control.
Stored Settings:
Up to 9 complete instrument set-ups may be stored and recalled from
non-volatile memory.
Size:
260(W) x 88(H) x 235(D)
Weight: 2kg.
(4.5lb.)
Power:
110-120V AC or 220V-240V AC ±10%, 50/60Hz,
adjustable internally; 35VA max. Installation Category II.
Operating Range:
+5°C to 40°C, 20-80% RH.
Storage Range:
-20°C to + 60°C.
Environmental:
Indoor use at altitudes up to 2000m, Pollution Degree 2.
Safety & EMC:
Complies with EN61010-1 & EN61326.
5