Calibration
Refer to the General section for dismantling instructions and safety precautions. Normal calibration
is done without opening the instrument.
Allow 10 minute warm-up before commencing calibration.
Refer to the User Manual for detailed operation of these power supplies.
Equipment Required
A 5.5 digit multimeter with better than 0.02% accuracy on dc volts and better than 0.12% accuracy
on dc current (to 5A); alternatively use a precision shunt for current measurement. An electronic
load set to approximately 2.7A or a 1.8
Ω
15W resistor.
Calibration
To enter calibration press
SHIFT,
#, 99. If the instrument is a long way out of calibration or has
been repaired, the default calibration values should be loaded first by pressing SHIFT #, 91.
Pressing ESCAPE at anytime will abort the calibration procedure and revert to the stored
calibration values.
Having entered calibration mode, follow the table below. To increment to the next step press OK.
The calibration
must
be done in sequence. At the end of calibration press OK to store the
calibration. The instrument automatically sets the range and settings at each step.
Use the spin-wheel to adjust calibration; holding down the JOG SET OFF key will give x100
increments to speed up calibration. Pressing SHIFT will show the calibration value.
On triple units calibrate both output 1 and output 2.
There are two internal adjustments which are factory set and will normally only require adjustment
if a component has been changed in the associated area. If this is so proceed as follows. VR1 and
VR2 are both located near the top of the control pcb. They are accessible from both sides of the
board. Turn them both fully clockwise as viewed from the component side, fully anti-clockwise as
viewed from the solderside. Adjustment is done with the output set to off. Connect an ammeter
across the output terminals and adjust VR1 for +1mA. Remove ammeter and connect voltmeter
and adjust for VR2 –10mV. (minus or negative 10mV).
Calibration steps 17 to 22 on output 1 are for the Auxiliary output.
QL355
Step/Range
Settings
Adjust for:-
Load
Detail
1, 35V 3A
0.01V 3A
10mV ± 0.5mV
DVM
o/p 0V
2, 35V 3A
0.01V 3A
0.010V on display
DVM
V readback zero
3, 35V 3A
35V 3A
35V ± 0.5mV
DVM
o/p V span
4, 35V 3A
35V 3A
35.000 on display
DVM
V readback span
5, 35V 3A
0.01V 3A
10mV ± 0.5mV
DVM
o/p 0V
6, 35V 3A
35V 3A
35V ± 0.5mV
DVM
o/p V span
7, 35V 3A
2V 0.001A
1mA ± 0.5mA
milli-ammeter
o/p A offset
8, 35V 3A
2V 0.001A
0.001 on display
milli-ammeter
A readback zero
9, 35V 500mA
2V 0.1mA
0.1mA ± 0.05mA
milli-ammeter
o/p mA offset
10, 35V 500mA
2V 0.1mA
flashing 0.0/0.1 on display
milli-ammeter
mA readback zero
change load
11, 15V 5A
2V 4A
4A ± 0.5mA
ammeter
o/p A span
12, 15V 5A
2V 4A
4.000 on display
ammeter
A readback span
13, 15V 5A
2V 4.1A
4.100 on display
ammeter
A readback 13bit
Continued….
Содержание QL II Series
Страница 1: ...QL Series II Precision Power Supplies Service Manual Book Part Number 48511 1740 Issue 1...
Страница 32: ...31 Block Diagram...
Страница 33: ...32 Interconnection Diagram SINGLE PSU TRIPLE PSU...
Страница 34: ...33 Component Layouts CONTROL PCB...
Страница 35: ...34 POWER PCB...
Страница 36: ...35 MAIN TERMINAL PCB AUXILIARY POWER PCB TRIPLE ONLY...
Страница 37: ...36 INTERFACE PCB...
Страница 38: ...37 KEYBOARD DISPLAY PCB SINGLE...
Страница 39: ...38 KEYBOARD DISPLAY PCB TRIPLE...
Страница 40: ...39 Circuit Diagrams AC SUPPLY CONNECTION PCB MAIN OUTPUT FRONT PANEL TERMINAL PCB...
Страница 41: ...QL SERIES II POWER PCB Sheet 1 of 2...
Страница 42: ...QL SERIES II POWER PCB Sheet 2 of 2...
Страница 43: ...DAC ADC Sample Hold ccts error logic Control Pcb Sheet 1 of 2...
Страница 44: ...Voltage Current Control OVP Sense fault trip Control Pcb Sheet 2of 2...
Страница 45: ...QL SERIES II AUXILIARY PCB...
Страница 46: ...QL SERIES II INTERFACE PCB Sheet 1 of 2...
Страница 47: ...QL SERIES II INTERFACE PCB Sheet 2 of 2...