Page 30
5
File Storage, File Types and USB Connections
The instrument incorporates a large Flash based disk drive for the storage of a variety of
different file types including Traces, Screen Images and Instrument Set-ups. When the
upgrade pack (U01) is fitted it may also include Logging Files, Limit Pattern Tables and
Amplitude Compensation Tables.
The instrument filing system provides a consistent method for storing, recalling, renaming,
deleting and copying files. Files can be transferred to and from a personal computer for
backup, analysis, printing and documentation purposes.
Whenever files are being accessed, the front panel LED marked Disk flashes. Depending upon
the file operation in progress, there may be a significant delay before further key operations are
possible.
5.1 Storing Traces or Screens
Storing of Traces or Screen Images is controlled from
the menu function:
Traces/Markers > Traces/Stores > Store.
(as described in section
This function enables a trace or screen image to be
stored as a file within the internal memory. (N.B.
external memory cannot be used to store files
directly, although files can be copied to external
memory later).
The menu keys are replaced by a control screen as
shown.
The top row of keys sets what will be stored: the Live
Trace, the View Trace, or a Screen Image. The
currently selected file type is shown in the green box
above the keys.
See section 5.7 for an explanation of Trace files and
Screen Image files.
There are two methods for saving to a file: Quick Save and Save As.
stores the trace or screen under an automatically incrementing name (e.g.
TRACE001 etc.) as set by the Default name as described below in section 5.1.1.1.
The key can be pressed repeatedly to store traces/images as they appear. The next
name that will be used is shown in yellow within the area above the keys.
halts the trace update (causing the sweep
status key to go to Pause) and enters the
next screen as shown.
The auto-allocated file name is shown (based upon
the current default file name), and pressing Save will
store the file under that name. However, the option
also exist to change the name before saving.
Pressing Change Name brings up an alpha-numeric
keypad which can be used to enter any name of up to
eight characters that the user may choose.
See section 5.6.1.