5-7 Page7 TIMING OUT setting
For more information about the TIMING OUT signal, see page 54.
1.
: SETUP channel selection
Same as Page 1.
2.
: TIMING OUT mode set
DIS/GATE/P.200ns/P.10us/P.100us/P.1ms
Manual/M.1us/M.10us/M.100us/M.1ms
default: P.1ms
Select
the type of
when outputting
TIMING OUT
signal
.
DIS
: Signal output
disabled.
GATE
: GATE mode output. Signal out between START and STOP.
P.200ns~1ms : Pulse output
of
200ns, 10us, 100us, 1ms.
Manual
: Manual output.
M.1u
s~1ms
: Outputs a pulse with a manually set width once.
3.
: TIMING OUT Start Position
Setting range:-2,147,483,647 ~ +2,147,483,647
Initial value: +1000
Set the starting point of the interval to perform TIMING OUT signal output of the
above.
4.
: TIMING OUT Stop Position
Setting range: -2,147,483,647 ~ +2,147,483,647
Initial value: +10000
Set the end point of the interval to perform TIMING OUT signal output of the above.
5.
: TIMING OUT Interval Pulse
Setting range: 0 ~ +2,147,483,647
Initial value: 200
When TIMING OUT signal is a pulse, you can set the pulse interval in TIMING
OUT output period.