5.  Press 

ESC

 twice to return to the Main Screen. 

 

UTILITY MENU

 

Pump Speed 

099 

Data Logging 

ON 

Location 

120 

 

 

Collecting Samples 

1.  Press 

RUN

 using the Left Soft Key. This turns on 

the pump and the particle counter runs for 10 cycles 
that are one minute long with 30 second hold times 
in-between. 

2.  Use the Buffer Menu to review the data.  

Software 

1.  The TrakPro

 Lite Secure Data Transfer utility and 

user manuals come on a CD that was packaged with 
your particle counter.  

2.  To install the communications software and drivers, 

insert the CD into your computer drive and follow 
the on-screen instructions.  

3.  See the 

TrakPro

 Lite Secure 

Software User’s 

Guide (

P/N 6004404) on CD (P/N 7001901) for 

installation instructions. 

 

N O T E

 

Make sure the particle counter is connected before 
you run the software. 

Checking Flow 

1.  Periodically check the inlet flow of the instrument 

with an external flow meter standard. 

2.  Connect a flowmeter standard to the inlet and turn 

on the instrument.  

3.  From the main menu, select the Utility menu 

4.  Select the Pump Speed and move the speed setting 

up or down as needed to match the flow standard at 
2.83 L/min (0.1 cfm). 

5.  When the flow is set, select 

ESC

 to go back to the 

main menu. 

6.  Disconnect the flowmeter. 
 

Содержание Aerotrak 9303

Страница 1: ... visible light Exposure to this light may cause blindness DO NOT remove any parts from the particle counter unless you are specifically told to do so in this manual DO NOT remove the housing or covers There are no user serviceable components inside the housing The use of controls adjustments or procedures other than those specified in this manual may result in exposure to hazardous optical radiati...

Страница 2: ... ordered accessories or spare parts Power 1 The battery is shipped separately and has to be installed before use 2 Remove the battery cover on the back side of the handle with a screwdriver and connect the battery to the socket Replace the battery cover Do not over tighten the screw 3 Fully charge the battery before use 4 Select the appropriate country specific power plug for the AC power adapter ...

Страница 3: ... the Left and Right Soft Keys under and to cycle between values 2 If the setting has more than a few possible values then the Left and Right Soft Keys are not available Instead select Enter to display a screen where the value can be edited with the Up Down keys as well as the Left and Right Soft Keys Use ESC to save the new value and return to the previous menu Set up Time Date 1 Select MENU to go...

Страница 4: ...te Secure Data Transfer utility and user manuals come on a CD that was packaged with your particle counter 2 To install the communications software and drivers insert the CD into your computer drive and follow the on screen instructions 3 See the TrakPro Lite Secure Software User s Guide P N 6004404 on CD P N 7001901 for installation instructions N O T E Make sure the particle counter is connected...

Отзывы: