3.  Follow the onscreen instructions to find the correct 

firmware upgrade. 

4.  Check firmware version on boot-up. Verify that 

firmware version is 1.26 or greater for 3787 or 
3788, and 2.15 or greater for 3772, 3775, or 3776. 

Operating the Electrostatic 
Classifier 

You can operate the Electrostatic Classifier in three 
modes:  

Classifier 
Mode 

Use Classifier tab. No CPC or computer 
required. 3082 acts as a size selector. 

 

SMPS 
Mode 

Use SMPS tab. CPC required. Set up, 
run, and log SMPS scans from display. 

 

External 
Control 
Mode 

Use Aerosol Instrument Manager 
software. CPC required. Set up, run, and 
log SMPS scans from a computer. 

 

Setup Tab 

Screens in the 

Setup

 tab provide options to set 

properties, view device details, set logging and 
communication settings, and calibrate instrument 
flows. 

Full details of the display options and modes are given 
in the Operation and Service Manual P/N 6006760.  

Title Bar 

The Title Bar indicates date and time, scan status, and 
errors. It also contains icons for toggling data logging 
and size distribution corrections. 

Blue 

Indicates the instrument is operating as 
expected. 

 

Yellow 

Indicates a warning condition. Touch the 
wrench to see a description of the warning 
condition. 

 

Red 

Indicates an error condition. When the title bar 
is red, the wrench icon flashes. Touch the 
wrench to see a description of the error 
condition.  

 

Setting Up a Basic SMPS Scan 

Although the correct scan settings for your application 
depend on the size range you are interested in and the 
type of measurement you are trying to make, the steps 
below can provide a starting point. 

1.  Follow steps in this QuickStart guide for connecting a 

CPC and neutralizer to the 3082. Verify that they are 
listed on the Classifier tab. 

2.  Press the 

 button to enable sheath flow. Set 

sheath flow to approximately 10X the aerosol flow of 
your CPC. 

3.  Navigate to the SMPS tab and press 

 to access 

scan properties.  

4.  On the Scan tab, set scan time (i.e., 30 seconds). On 

the Schedule tab, set Scans per Sample (i.e., 1) and 
Number of Samples (i.e., 1). 

5.  Press 

OK

 to return to SMPS tab. Press

 

 to start. 

Size distribution data and statistics are displayed. 

Moving and Shipping 

1.  Power off the classifier and disconnect the power 

cord from the power supply. 

2.  If using a DMA, disconnect all tubing and cap the 

fittings, disconnect the high-voltage supply, and then 
depress the lever on the classifier base plate to 
release the DMA.  

3.  If installed, remove the Neutralizer.  

4.  If installed, remove the optional lead shielding. 

5.  When shipping, place the classifier in the original 

packing materials. The classifier should be placed in 
the smaller of the two supplied boxes and anchored 
with the foam inserts. The box should be taped shut 
and then placed in the larger of the two boxes, 
protected at the corners by foam inserts. 

Note

:  If your impactors require calibration, return 

them along with the classifier to TSI.  

Содержание 3082

Страница 1: ...structs you to do so Do not remove any part of the instrument housing while the instrument is powered on Unpacking Carefully unpack the classifier Verify that all items in the packing list see below are present Contact TSI if items are missing or broken Complete details of the 3082 Accessory Kit DMA Upgrade Kits and optional accessories are given in the Operation and Service Manual Retain the pack...

Страница 2: ...ved in Step 1 Installing Model 3088 Neutralizer The 3088 Advanced Aerosol Neutralizer has the neutralization capacity of the 3077A neutralizer but has an x ray source instead of a Kr 85 source which can be turned on and off To install 1 Remove the flag assembly 2 screws from the top of the classifier and store in Accessory kit 2 Remove the black vinyl end caps from the 3088 and install the 3088 in...

Страница 3: ...igh voltage connector into the high voltage port on the classifier Align the flattened side of the connector with the flattened side of the port Note For alternative installation of the 1nm DMA to minimize diffusion losses between the Model 3777 Nano Enhancer and the 1nm DMA refer to the Operation and Service Manual Connecting Classifier to Detector The Model 3082 supports communication to a 3772 ...

Страница 4: ...ench to see a description of the error condition Setting Up a Basic SMPS Scan Although the correct scan settings for your application depend on the size range you are interested in and the type of measurement you are trying to make the steps below can provide a starting point 1 Follow steps in this QuickStart guide for connecting a CPC and neutralizer to the 3082 Verify that they are listed on the...

Отзывы: