Applications
Model 3079A
5-5
The following table gives an overview of the expected particle size depending on salt solution
concentration and primary droplet size.
Concentration of Salt Solution
Residual Size for
1 µm Droplet
Residual Size for
5 µm Droplet
5.000%
50000 ppm
0.368 µm
1.842 µm
1.000%
10000 ppm
0.215 µm
1.077 µm
0.500%
5000 ppm
0.171 µm
0.855 µm
0.100%
1000 ppm
0.100 µm
0.500 µm
0.020%
200 ppm
0.058 µm
0.292 µm
0.010%
100 ppm
0.046 µm
0.232 µm
0.005%
50 ppm
0.037 µm
0.184 µm
0.001%
10 ppm
0.022 µm
0.108 µm
Applications
Generation of very fine aerosols with particle diameters <100 nm
Filter testing
Содержание 3079A