2. Temperature Calibration Adjust
The temperature calibration adjustment adds or subtracts a constant number
of degrees Fahrenheit (°F) or degrees Celsius (°C) from the displayed
reading. The adjustment range is ±10.0°F or ±5.5°C. Temperature
calibration adjustments can be done in either the °F or °C scale depending
on temperature scale selected.
Turn the instrument power off and remove batteries to gain access to DIP
switches. Set DIP switches 7 and 5 to ON position. Replace batteries and
turn the instrument on. The instrument is in the Temperature Calibration
Adjust Mode and will display 0.0 (or any previously set value) and the
temperature scale (°F or °C). Using
▲
or
▼
keys set the display to a desired
offset value and press
ENTER
key. The instrument will display
dOnE.
At
this point turn the instrument off, remove batteries and reset DIP switches 5
and 7 to OFF position. Replace batteries and the battery cover. Turn the
instrument on. During the start-up routine the display will show briefly the
newly selected temperature adjustment offset (the instrument does not
display the offset value if it is set to 0.0).
To return to the original factory calibration settings, set the percentage
adjustment factor to 0.0% following the steps described above.